WEKO3
アイテム
A design of forceps-type coincidence radiation detector for intraoperative LN diagnosis: clinical impact estimated from LNs data of 20 esophageal cancer patients
https://repo.qst.go.jp/records/84998
https://repo.qst.go.jp/records/84998c9224b1b-83b8-40cc-87bd-2f84d0ec1011
Item type | 学術雑誌論文 / Journal Article(1) | |||||
---|---|---|---|---|---|---|
公開日 | 2021-12-08 | |||||
タイトル | ||||||
タイトル | A design of forceps-type coincidence radiation detector for intraoperative LN diagnosis: clinical impact estimated from LNs data of 20 esophageal cancer patients | |||||
言語 | ||||||
言語 | eng | |||||
資源タイプ | ||||||
資源タイプ識別子 | http://purl.org/coar/resource_type/c_6501 | |||||
資源タイプ | journal article | |||||
アクセス権 | ||||||
アクセス権 | metadata only access | |||||
アクセス権URI | http://purl.org/coar/access_right/c_14cb | |||||
著者 |
Miwako, Takahashi
× Miwako, Takahashi× Yoshimura, Shuntaro× Sodai, Takyu× Aikou, Susumu× Okumura, Yasuhiro× Yagi, Koichi× Fukayama, Masashi× Momose, Toshimitsu× Seto, Yasuyuki× Taiga, Yamaya× Miwako, Takahashi× Sodai, Takyu× Seto, Yasuyuki× Taiga, Yamaya |
|||||
抄録 | ||||||
内容記述タイプ | Abstract | |||||
内容記述 | Purpose To reduce postoperative complications, intraoperative lymph node (LN) diagnosis with 18F-fluoro-2-deoxy-D-glucose (FDG) is expected to optimize the extent of LN dissection, leading to less invasive surgery. However, such a diagnostic device has not yet been realized. We proposed the concept of coincidence detection wherein a pair of scintillation crystals formed the head of the forceps. To estimate the clinical impact of this detector, we determined the cut-off value using FDG as a marker for intraoperative LN diagnosis in patients with esophageal cancer, the specifications needed for the detector, and its feasibility using numerical simulation. Methods We investigated the dataset including pathological diagnosis and radioactivity of 1073 LNs resected from 20 patients who underwent FDG-positron emission tomography followed by surgery for esophageal cancer on the same day. The specifications for the detector were determined assuming that it should measure 100 counts (less than 10% statistical error) or more within the intraoperative measurement time of 30 s. The detector sensitivity was estimated using GEANT4 simulation and the expected diagnostic ability was calculated. Results The cut-off value was 620 Bq for intraoperative LN diagnosis. The simulation study showed that the detector had a radiation detection sensitivity of 0.96%, which was better than the estimated specification needed for the detector. Among the 1035 non-metastatic LNs, 815 were below the cut-off value. Conclusion The forceps-type coincidence detector can provide sufficient sensitivity for intraoperative LN diagnosis. Approximately 80% of the prophylactic LN dissections in esophageal cancer can be avoided using this detector. |
|||||
書誌情報 |
Annals of Nuclear Medicine 巻 36, p. 285-292, 発行日 2021-11 |
|||||
出版者 | ||||||
出版者 | Springer Nature | |||||
ISSN | ||||||
収録物識別子タイプ | ISSN | |||||
収録物識別子 | 0914-7187 | |||||
DOI | ||||||
識別子タイプ | DOI | |||||
関連識別子 | 10.1007/s12149-021-01701-9 | |||||
関連サイト | ||||||
識別子タイプ | DOI | |||||
関連識別子 | https://doi.org/10.1007/s12149-021-01701-9 |