Experimental evaluation of scattered X-ray spectra due to X-ray therapeutic and diagnosis equipment for eye lens dosimetry of medical staff
https://repo.qst.go.jp/records/84999
https://repo.qst.go.jp/records/849992656582f-9caf-410b-a3e2-96a47e6ddf09
Item type | 学術雑誌論文 / Journal Article(1) | |||||
---|---|---|---|---|---|---|
公開日 | 2022-02-22 | |||||
タイトル | ||||||
タイトル | Experimental evaluation of scattered X-ray spectra due to X-ray therapeutic and diagnosis equipment for eye lens dosimetry of medical staff | |||||
言語 | ||||||
言語 | eng | |||||
資源タイプ | ||||||
資源タイプ識別子 | http://purl.org/coar/resource_type/c_6501 | |||||
資源タイプ | journal article | |||||
アクセス権 | ||||||
アクセス権 | metadata only access | |||||
アクセス権URI | http://purl.org/coar/access_right/c_14cb | |||||
著者 |
Munehiko, Kowatari
× Munehiko, Kowatari× Nagamoto, Keisuke× Nakagami, Koich× Tanimura, Yoshihiko× Takashi, Moritake× Kunugita, Naoki× Munehiko, Kowatari× Koichi, Nakagami× Takashi, Moritake× Naoki, Kunugita |
|||||
抄録 | ||||||
内容記述タイプ | Abstract | |||||
内容記述 | We have been investigating to properly evaluate personal dose equivalent for the lens of the eye, Hp(3) of highly exposed medical staffs. One of the key parameters is the mean X-ray energy which medical staff may receive around their eyes, because the response of the dosimeter is strongly dependent on incident X-ray energy. However, there is a few studies which experimentally investigated scattered X-rays that medical staff actually encounters during their procedures. The scattered X-rays from X-ray diagnosis and therapeutic equipment was measured using the CdTeZn spectrometer. Derived results indicated that the mean X-ray energies ranged between 40 keV and 60 keV. The measurement results also support that the reduction effect on the eye lens dose due to the protection glasses or ceiling-suspended shield was experimentally evaluated from the pulse height spectra. | |||||
書誌情報 |
Journal of Radiation Protection and Research 巻 47, 号 1, p. 39-49, 発行日 2022-02 |
|||||
ISSN | ||||||
収録物識別子タイプ | ISSN | |||||
収録物識別子 | 2508-1888 | |||||
DOI | ||||||
識別子タイプ | DOI | |||||
関連識別子 | 10.14407/jrpr.2021.00220 | |||||
関連サイト | ||||||
識別子タイプ | DOI | |||||
関連識別子 | https://doi.org/10.14407/jrpr.2021.00220 |