WEKO3
アイテム
Report of AAPM Task Group 290: Respiratory motion management for particle therapy
https://repo.qst.go.jp/records/84950
https://repo.qst.go.jp/records/8495047322b38-c030-4052-addf-dd4206edbf47
Item type | 学術雑誌論文 / Journal Article(1) | |||||
---|---|---|---|---|---|---|
公開日 | 2022-01-27 | |||||
タイトル | ||||||
タイトル | Report of AAPM Task Group 290: Respiratory motion management for particle therapy | |||||
言語 | ||||||
言語 | eng | |||||
資源タイプ | ||||||
資源タイプ識別子 | http://purl.org/coar/resource_type/c_6501 | |||||
資源タイプ | journal article | |||||
アクセス権 | ||||||
アクセス権 | metadata only access | |||||
アクセス権URI | http://purl.org/coar/access_right/c_14cb | |||||
著者 |
Li, Heng
× Li, Heng× Dong, Lei× Christoph Bert× Joe Chang× Stella Flampouri× Kyung-Wook Jee× Liyong Lin× Moyers, Michael× Shinichiro, Mori× Joerg Rottmann× Erik Tryggestad× Vedam, Sastry× Shinichiro, Mori |
|||||
抄録 | ||||||
内容記述タイプ | Abstract | |||||
内容記述 | Dose uncertainty induced by respiratory motion remains a major concern for treating thoracic and abdominal lesions using particle beams. This Task Group report reviews the impact of tumor motion and dosimetric considerations in particle radiotherapy, current motion-management techniques, and limitations for different particle-beam delivery modes (i.e., passive scattering, uniform scanning, and pencil-beam scanning). Furthermore, the report provides guidance and risk analysis for quality assurance of the motion-management procedures to ensure consistency and accuracy, and discusses future development and emerging motion-management strategies. This report supplements previously published AAPM report TG76, and considers aspects of motion management that are crucial to the accurate and safe delivery of particle-beam therapy. To that end, this report produces general recommendations for commissioning and facility-specific dosimetric characterization, motion assessment, treatment planning, active and passive motion-management techniques, image guidance and related decision-making, monitoring throughout therapy, and recommendations for vendors. Key among these recommendations are that: 1) facilities should perform thorough planning studies (using retrospective data) and develop standard operating procedures that address all aspects of therapy for any treatment site involving respiratory motion; 2) a risk-based methodology should be adopted for quality management and ongoing process improvement. | |||||
書誌情報 |
Medical Physics 巻 49, 号 4, p. e50-e81, 発行日 2022-01 |
|||||
出版者 | ||||||
出版者 | Wiley | |||||
ISSN | ||||||
収録物識別子タイプ | ISSN | |||||
収録物識別子 | 0094-2405 | |||||
DOI | ||||||
識別子タイプ | DOI | |||||
関連識別子 | 10.1002/mp.15470 | |||||
関連サイト | ||||||
識別子タイプ | URI | |||||
関連識別子 | https://aapm.onlinelibrary.wiley.com/doi/10.1002/mp.15470 |