WEKO3
アイテム
Performance evaluation of a staggered three-layer DOI PET detector using a 1 mm LYSO pitch with PETsys TOFPET2 ASIC: comparison of HAMAMATSU and KETEK SiPMs
https://repo.qst.go.jp/records/84044
https://repo.qst.go.jp/records/84044a9c52612-d560-4d50-bbb5-90b9c7b13a92
Item type | 学術雑誌論文 / Journal Article(1) | |||||
---|---|---|---|---|---|---|
公開日 | 2021-07-19 | |||||
タイトル | ||||||
タイトル | Performance evaluation of a staggered three-layer DOI PET detector using a 1 mm LYSO pitch with PETsys TOFPET2 ASIC: comparison of HAMAMATSU and KETEK SiPMs | |||||
言語 | ||||||
言語 | eng | |||||
資源タイプ | ||||||
資源タイプ識別子 | http://purl.org/coar/resource_type/c_6501 | |||||
資源タイプ | journal article | |||||
アクセス権 | ||||||
アクセス権 | metadata only access | |||||
アクセス権URI | http://purl.org/coar/access_right/c_14cb | |||||
著者 |
Binder, Tim
× Binder, Tim× Kang, Hangyu× Munetaka, Nitta× Schneider, Florian× Taiga, Yamaya× Parodi, Katia× Wiest, Florian× G Thirolf, Peter× Kang, Hangyu× Munetaka, Nitta× Taiga, Yamaya× Parodi, Katia |
|||||
抄録 | ||||||
内容記述タイプ | Abstract | |||||
内容記述 | In this study, we propose a staggered three-layer depth-of-interaction (DOI) detector with a 1 mm crystal pitch and 19.8 mm total crystal thickness for a high-resolution and high-sensitivity small animal in-beam PET scanner. A three-layered stacked LYSO scintillation array (0.9 × 0.9 × 6.6 mm3 crystals, 23 × 22 mm2 surface area)read out by a SiPM array (8 × 8 channels, 3 × 3 mm2 active area/channel and 50 μm microcell size)with data acquisition, signal processing and digitization performed using the PETsys Electronics Evaluations kit(based on the TOFPET v2c ASIC) builds a DOI LYSO detector block. The performance of the DOI detector was evaluated in terms of crystal resolvability, energy resolution, and coincidence resolving time (CRT). A comparative performance evaluation of the staggered three-layer LYSO block was conducted with two different SiPM arrays from KETEK and HAMAMATSU. 100% (KETEK) and 99.8% (HAMAMATSU) of the crystals were identified, by using a flood irradiation the front- and back-side. The average energy resolutions for the 1st, 2nd, and 3rd layers were 16.5 (±2.3)%, 20.9(±4.0)%, and 32.7 (±21.0)% (KETEK) and 19.3 (±3.5)%, 21.2 (±4.1)%, and 26.6 (±10.3)% (HAMAMATSU)for the used SiPM arrays. The measured CRTs(FWHM)for the 1st, 2nd, and 3rd layers were 532 (±111) ps, 463 (±108) ps, and 447 (±111) ps (KETEK) and 402 (±46) ps, 392 (±54) ps, and 408 (±196) ps(HAMAMATSU). In conclusion, the performance of the staggered three-layer DOI detector with 1 mm LYSO pitch and 19.8 mm total crystal thickness was fully characterized. The feasibility of a highly performing readout of a high resolution DOI PET detector via SiPM arrays from KETEK and HAMAMATSU employing the PETsys TOFPET v2c ASIC could be demonstrated. | |||||
書誌情報 |
Physics in Medicine and Biology 巻 66, 号 12, p. 125016, 発行日 2021-06 |
|||||
出版者 | ||||||
出版者 | IOP Publishing | |||||
ISSN | ||||||
収録物識別子タイプ | ISSN | |||||
収録物識別子 | 0031-9155 | |||||
DOI | ||||||
識別子タイプ | DOI | |||||
関連識別子 | 10.1088/1361-6560/abfbf3 | |||||
関連サイト | ||||||
識別子タイプ | DOI | |||||
関連識別子 | https://doi.org/10.1088/1361-6560/abfbf3 |