WEKO3
アイテム
Alternating-phase-focused linac for an injector of medical synchrotrons
https://repo.qst.go.jp/records/67832
https://repo.qst.go.jp/records/678326c88eeea-6f50-4a51-bb64-2b0df4aaaa90
Item type | 会議発表用資料 / Presentation(1) | |||||
---|---|---|---|---|---|---|
公開日 | 2004-07-12 | |||||
タイトル | ||||||
タイトル | Alternating-phase-focused linac for an injector of medical synchrotrons | |||||
資源タイプ | ||||||
資源タイプ識別子 | http://purl.org/coar/resource_type/c_c94f | |||||
資源タイプ | conference object | |||||
アクセス権 | ||||||
アクセス権 | metadata only access | |||||
アクセス権URI | http://purl.org/coar/access_right/c_14cb | |||||
著者 |
Iwata, Yoshiyuki
× Iwata, Yoshiyuki× 岩田 佳之 |
|||||
抄録 | ||||||
内容記述タイプ | Abstract | |||||
内容記述 | A preliminary design of a compact heavy-ion injector for medical synchrotrons was presented. The compact injector consists of the Radio-Frequency-Quadrupole (RFQ) linac and Interdigital H-mode (IH) Drift-Tube-Linac (DTL). For the IH-DTL, Alternating-Phase-Focusing (APF) method will be applied. The electric field distribution of the IH cavity was calculated with a three-dimensional electromagnetic field solver and used for the simulations on the beam dynamics. By performing iterative calculations of the electric field and beam dynamics, the satisfactory emittance and acceptance were obtained. The compact injector will be constructed, and their beam test will be performed by the end of FY2005. | |||||
会議概要(会議名, 開催地, 会期, 主催者等) | ||||||
内容記述タイプ | Other | |||||
内容記述 | 9th European Particle Accelerator Conference (EPAC04) | |||||
発表年月日 | ||||||
日付 | 2004-07-09 | |||||
日付タイプ | Issued |