Introduction of PIXE anallysis system in National Institute of Radiological Sciences

https://repo.qst.go.jp/records/59906
https://repo.qst.go.jp/records/59906
dfaa37c5-6cb2-485c-b7f5-80781cba82c9
Item type 会議発表用資料 / Presentation(1)
公開日 2003-08-26
タイトル
タイトル Introduction of PIXE anallysis system in National Institute of Radiological Sciences
言語
言語 eng
資源タイプ
資源タイプ識別子 http://purl.org/coar/resource_type/c_c94f
資源タイプ conference object
アクセス権
アクセス権 metadata only access
アクセス権URI http://purl.org/coar/access_right/c_14cb
著者 Yukawa, Masae

× Yukawa, Masae

WEKO 595948

Yukawa, Masae

Search repository
Imaseki, Hitoshi

× Imaseki, Hitoshi

WEKO 595949

Imaseki, Hitoshi

Search repository
湯川 雅枝

× 湯川 雅枝

WEKO 595950

en 湯川 雅枝

Search repository
今関 等

× 今関 等

WEKO 595951

en 今関 等

Search repository
抄録
内容記述タイプ Abstract
内容記述 An electrostatic accelerator, Tandetron (Model 4117MC, High Voltage
Engineering Europe Co.) was installed in National Institute of Radiological
Sciences (NIRS) for PIXE (Particle Induced X-ray Emission) analysis in
1999. The accelerating voltage is 0.4 to 1.7MV, and the maximum beam
current is 5microA at 3.4MeV. This system has three beam ports for
different types of PIXE analysis, conventional, micro-beam and in-air.
Normal PIXE line has two types of X-ray detecting device, Si (Li) and
CdZnTe detectors detectable for elements from Na (Z=11) to U (Z=92).
Fifteen samples can semi automatically be measured at one time with the
optical beam size from 0.5 to 2.0 mm at 100 nA of the beam current. A
quadruple triplet magnet (Model OM2000, Oxford Micro beams, Ltd.) attached
to the other beam port produces a proton micro-beam of the square shape
less than 1micron x 1micron. Micro-beam
scanning PIXE analysis is carried out with this beam at 50pA current and
scanning area up to 2.0mm square. The in-air PIXE analysis is performed
using the third beam port.
In this paper, we introduce micro-PIXE scanning system showing the ability
to draw elemental maps in a spatial resolution of about 1 µm with
analyses of fish scale, fish otolith, crocus pollen and micro organisms.
Fine ring structure of a fish scale was observed using elemental mapping
with proton micro-beam scanning. Pollen was analyzed as one example of
single cell to demonstrate the elemental distribution. Fish scale and
otolith seem to be a kind of time recorder like tree rings. And the
elemental composition reflects the living condition. Therefore, fish scale
and otolith are possible to be used as a biological indicator.
会議概要(会議名, 開催地, 会期, 主催者等)
内容記述タイプ Other
内容記述 3rd International Symposium on Nuclear Analytical Chemistry (NAC-III)
発表年月日
日付 2001-06-14
日付タイプ Issued
戻る
0
views
See details
Views

Versions

Ver.1 2023-05-15 21:59:42.842880
Show All versions

Share

Mendeley Twitter Facebook Print Addthis

Cite as

エクスポート

OAI-PMH
  • OAI-PMH JPCOAR 2.0
  • OAI-PMH JPCOAR 1.0
  • OAI-PMH DublinCore
  • OAI-PMH DDI
Other Formats
  • JSON
  • BIBTEX

Confirm


Powered by WEKO3


Powered by WEKO3