Technical consultation to the IAEA regarding dose audit system using radiophotoluminescent glass dosimeter developed by NIRS

https://repo.qst.go.jp/records/55956
https://repo.qst.go.jp/records/55956
64d767e7-4be9-4f17-9deb-bfe3824f68a9
Item type 一般雑誌記事 / Article(1)
公開日 2015-08-18
タイトル
タイトル Technical consultation to the IAEA regarding dose audit system using radiophotoluminescent glass dosimeter developed by NIRS
言語
言語 eng
資源タイプ
資源タイプ識別子 http://purl.org/coar/resource_type/c_6501
資源タイプ article
アクセス権
アクセス権 metadata only access
アクセス権URI http://purl.org/coar/access_right/c_14cb
著者 水野, 秀之

× 水野, 秀之

WEKO 572582

水野, 秀之

Search repository
水野 秀之

× 水野 秀之

WEKO 572583

en 水野 秀之

Search repository
抄録
内容記述タイプ Abstract
内容記述 In Japan, postal dose audits have been performed on radiation
therapy units using radiophotoluminescent glass dosimeters
(RGDs), since 2007 [1]. NIRS developed this auditing system and
has supported its technical aspects continuously. For instance,
the application of RGDs to a non-reference condition beam such
as field size change or wedged field were studied by NIRS and
successfully introduced in the routine dose audit in 2010 [2] [3].
The IAEA / WHO have performed postal dose audits worldwide
since 1969 using thermoluminescent dosimeters (TLDs). Recently,
They have considered the replacement of TLDs by RGDs.
RGDs are made in Japan and offer some superior features: RGDs
are almost completely free from the fading effect and can be read
repeatedly. NIRS and IAEA have done an annual dosimetric intercomparison
between RGDs and TLDs since 2008 (Fig.1). The outputs
of both dosimeters agreed within ±1% (Table1). In 2013,
IAEA installed a new RGD reader from Japan and started commissioning
of the RGD system. IAEA asked NIRS to support this
work and I stayed at the IAEA Dosimetry Laboratory for 1 year
from August 2013 as a consultant on the RGD system commissioning.
書誌情報 National Institute of Radiological Sciences Annual Report 2014

巻 2014, p. 100-101, 発行日 2015-08
ISSN
収録物識別子タイプ ISSN
収録物識別子 0439-5956
戻る
0
views
See details
Views

Versions

Ver.1 2023-05-15 22:43:27.336910
Show All versions

Share

Mendeley Twitter Facebook Print Addthis

Cite as

エクスポート

OAI-PMH
  • OAI-PMH JPCOAR 2.0
  • OAI-PMH JPCOAR 1.0
  • OAI-PMH DublinCore
  • OAI-PMH DDI
Other Formats
  • JSON
  • BIBTEX

Confirm


Powered by WEKO3


Powered by WEKO3