Ultra-trace determination of Am-241 by SF-ICP-MS combined with a high efficiency sample introduction system
https://repo.qst.go.jp/records/54797
https://repo.qst.go.jp/records/5479779c133ea-0b84-45d8-ada8-75deada8ba75
Item type | 会議発表論文 / Conference Paper(1) | |||||
---|---|---|---|---|---|---|
公開日 | 2017-07-13 | |||||
タイトル | ||||||
タイトル | Ultra-trace determination of Am-241 by SF-ICP-MS combined with a high efficiency sample introduction system | |||||
言語 | ||||||
言語 | eng | |||||
資源タイプ | ||||||
資源タイプ識別子 | http://purl.org/coar/resource_type/c_5794 | |||||
資源タイプ | conference paper | |||||
アクセス権 | ||||||
アクセス権 | metadata only access | |||||
アクセス権URI | http://purl.org/coar/access_right/c_14cb | |||||
著者 |
Zheng, Jian
× Zheng, Jian× Wang, Zhongtang× Tagami, Keiko× Uchida, Shigeo× 鄭 建× 王 鍾堂× 田上 恵子× 内田 滋夫 |
|||||
抄録 | ||||||
内容記述タイプ | Abstract | |||||
内容記述 | In this study, a detail evaluation is made of the new sector-field ICP-MS (Element XR) on its analytical potential for ultra-trace determination of 241Am. In order to enhance the sensitivity for 241Am analysis, a high efficiency sample introduction system (Aridus II) was combined with the Element XR. Analytical parameters, such as the RF power of SF-ICP-MS, flow rates of Ar sweep gas and N2 gas of the Aridus system were carefully optimized. A detection limit at attogram (10-18 g) level was obtained, which is comparable or even better than that of AMS, indicating a significant advance of ICP-MS for ultra-trace 241Am measurements. | |||||
書誌情報 |
Proceedings of the 17th Workshop on Environmental Radioactivity 巻 2016-8, p. 180-185, 発行日 2016-10 |