Dynamic splitting of Gaussian pencil beams in heterogeneity-correction algorithms for radiotherapy with heavy charged particles

https://repo.qst.go.jp/records/47578
https://repo.qst.go.jp/records/47578
1ff73d71-1f1c-4d4e-95ba-c9766e21c363
名前 / ファイル ライセンス アクション
c540e8b7a34d6454c8663fe983a861dc.pdf text.pdf (808.2 kB)
Item type 学術雑誌論文 / Journal Article(1)
公開日 2009-02-21
タイトル
タイトル Dynamic splitting of Gaussian pencil beams in heterogeneity-correction algorithms for radiotherapy with heavy charged particles
言語
言語 eng
資源タイプ
資源タイプ識別子 http://purl.org/coar/resource_type/c_6501
資源タイプ journal article
著者 Kanematsu, Nobuyuki

× Kanematsu, Nobuyuki

WEKO 773915

Kanematsu, Nobuyuki

Search repository
Komori, Masataka

× Komori, Masataka

WEKO 773916

Komori, Masataka

Search repository
Yonai, Shunsuke

× Yonai, Shunsuke

WEKO 773917

Yonai, Shunsuke

Search repository
Ishizaki, Azusa

× Ishizaki, Azusa

WEKO 773918

Ishizaki, Azusa

Search repository
Kanematsu, Nobuyuki

× Kanematsu, Nobuyuki

WEKO 773919

en Kanematsu, Nobuyuki

Search repository
Komori, Masataka

× Komori, Masataka

WEKO 773920

en Komori, Masataka

Search repository
Yonai, Shunsuke

× Yonai, Shunsuke

WEKO 773921

en Yonai, Shunsuke

Search repository
Ishizaki, Azusa

× Ishizaki, Azusa

WEKO 773922

en Ishizaki, Azusa

Search repository
抄録
内容記述タイプ Abstract
内容記述 The pencil-beam algorithm is valid only when elementary Gaussian beams are small enough compared to lateral heterogeneity of a medium, which is not always true in actual radiotherapy with protons and ions. This work addresses a solution for the problem. We found approximate self-similarity of Gaussian distributions, with which Gaussian beams can split into narrower and deflecting daughter beams when their sizes have overreached lateral heterogeneity in beam-transport calculation. The effectiveness was assessed in an carbon-ion beam experiment in the presence of steep range compensation, where the splitting calculation reproduced the detour effect amounting up to about 10% in dose or as large as the lateral particle disequilibrium effect. The efficiency was analyzed in calculations for carbon-ion and proton radiations with a heterogeneous phantom model, where the beam splitting increased computing times by factors of 4.7 and 3.2. The present method generally improves accuracy of the pencil-beam algorithm without severe inefficiency. It will therefore be useful for treatment planning and potentially other demanding applications.
書誌情報 Physics in Medicine and Biology

巻 54, 号 7, p. 2015-2027, 発行日 2009-04
出版者
出版者 IOP Publishing
ISSN
収録物識別子タイプ ISSN
収録物識別子 0031-9155
PubMed番号
識別子タイプ PMID
関連識別子 19287085
DOI
識別子タイプ DOI
関連識別子 10.1088/0031-9155/54/7/010
関連サイト
識別子タイプ DOI
関連識別子 https://doi.org/10.1088/0031-9155/54/7/010
戻る
0
views
See details
Views