ログイン
言語:

WEKO3

  • トップ
  • ランキング
To
lat lon distance
To

Field does not validate



インデックスリンク

インデックスツリー

メールアドレスを入力してください。

WEKO

One fine body…

WEKO

One fine body…

アイテム

  1. 原著論文

Field-size effect of physical doses in carbon-ion scanning using range shifter plates

https://repo.qst.go.jp/records/45669
https://repo.qst.go.jp/records/45669
ad123568-b8d7-4394-b0af-b00736594f24
Item type 学術雑誌論文 / Journal Article(1)
公開日 2009-11-10
タイトル
タイトル Field-size effect of physical doses in carbon-ion scanning using range shifter plates
言語
言語 eng
資源タイプ
資源タイプ識別子 http://purl.org/coar/resource_type/c_6501
資源タイプ journal article
アクセス権
アクセス権 metadata only access
アクセス権URI http://purl.org/coar/access_right/c_14cb
著者 Inaniwa, Taku

× Inaniwa, Taku

WEKO 453834

Inaniwa, Taku

Search repository
Furukawa, Takuji

× Furukawa, Takuji

WEKO 453835

Furukawa, Takuji

Search repository
Nagano, Ai

× Nagano, Ai

WEKO 453836

Nagano, Ai

Search repository
Satou, Shinji

× Satou, Shinji

WEKO 453837

Satou, Shinji

Search repository
Saotome, Naoya

× Saotome, Naoya

WEKO 453838

Saotome, Naoya

Search repository
Noda, Kouji

× Noda, Kouji

WEKO 453839

Noda, Kouji

Search repository
Kanai, Tatsuaki

× Kanai, Tatsuaki

WEKO 453840

Kanai, Tatsuaki

Search repository
稲庭 拓

× 稲庭 拓

WEKO 453841

en 稲庭 拓

Search repository
古川 卓司

× 古川 卓司

WEKO 453842

en 古川 卓司

Search repository
永野 あい

× 永野 あい

WEKO 453843

en 永野 あい

Search repository
佐藤 眞二

× 佐藤 眞二

WEKO 453844

en 佐藤 眞二

Search repository
早乙女 直也

× 早乙女 直也

WEKO 453845

en 早乙女 直也

Search repository
野田 耕司

× 野田 耕司

WEKO 453846

en 野田 耕司

Search repository
金井 達明

× 金井 達明

WEKO 453847

en 金井 達明

Search repository
抄録
内容記述タイプ Abstract
内容記述 A field-size effect of physical doses was studied in scanning irradiation with carbon ions. For the target volumes of 60*60*80, 40*40*80, and 20*20*80 mm^3, the doses along the beam axis within the spread-out Bragg peaks reduced to 99.4%, 98.2%, and 96.0% of the dose for the target of 80*80*80 mm^3, respectively. The present study revealed that the observed reductions can be compensated for by adopting the three-Gaussian form of lateral dose distributions for the pencil beam model used in the treatment planning system. The parameters describing the form were determined through the irradiation experiments making flat concentric squared frames with a scanned carbon beam. Since utilizing the three-Gaussian model in the dose optimization loop is at present time consuming, the correction for the field-size effect should be implemented as a "predicted-dose scaling factor". The validity of this correction method was confirmed through the irradiation of a target of 20*20*80 mm^3.
書誌情報 Medical Physics

巻 36, 号 7, p. 2889-2897, 発行日 2009-06
ISSN
収録物識別子タイプ ISSN
収録物識別子 0094-2405
DOI
識別子タイプ DOI
関連識別子 10.1118/1.3140586
戻る
0
views
See details
Views

Versions

Ver.1 2023-05-16 00:01:02.403282
Show All versions

Share

Mendeley Twitter Facebook Print Addthis

Cite as

エクスポート

OAI-PMH
  • OAI-PMH JPCOAR 2.0
  • OAI-PMH JPCOAR 1.0
  • OAI-PMH DublinCore
  • OAI-PMH DDI
Other Formats
  • JSON
  • BIBTEX

Confirm


Powered by WEKO3