@article{oai:repo.qst.go.jp:00045669, author = {Inaniwa, Taku and Furukawa, Takuji and Nagano, Ai and Satou, Shinji and Saotome, Naoya and Noda, Kouji and Kanai, Tatsuaki and 稲庭 拓 and 古川 卓司 and 永野 あい and 佐藤 眞二 and 早乙女 直也 and 野田 耕司 and 金井 達明}, issue = {7}, journal = {Medical Physics}, month = {Jun}, note = {A field-size effect of physical doses was studied in scanning irradiation with carbon ions. For the target volumes of 60*60*80, 40*40*80, and 20*20*80 mm^3, the doses along the beam axis within the spread-out Bragg peaks reduced to 99.4%, 98.2%, and 96.0% of the dose for the target of 80*80*80 mm^3, respectively. The present study revealed that the observed reductions can be compensated for by adopting the three-Gaussian form of lateral dose distributions for the pencil beam model used in the treatment planning system. The parameters describing the form were determined through the irradiation experiments making flat concentric squared frames with a scanned carbon beam. Since utilizing the three-Gaussian model in the dose optimization loop is at present time consuming, the correction for the field-size effect should be implemented as a "predicted-dose scaling factor". The validity of this correction method was confirmed through the irradiation of a target of 20*20*80 mm^3.}, pages = {2889--2897}, title = {Field-size effect of physical doses in carbon-ion scanning using range shifter plates}, volume = {36}, year = {2009} }