{"_buckets": {"deposit": "4a8fd90b-93b7-4a49-8e6c-6d25b0f3179d"}, "_deposit": {"created_by": 1, "id": "82943", "owners": [1], "pid": {"revision_id": 0, "type": "depid", "value": "82943"}, "status": "published"}, "_oai": {"id": "oai:repo.qst.go.jp:00082943", "sets": ["1"]}, "author_link": ["997033", "997032", "997035", "997031", "997029", "997034", "997030", "997028"], "item_8_biblio_info_7": {"attribute_name": "書誌情報", "attribute_value_mlt": [{"bibliographicIssueDates": {"bibliographicIssueDate": "2021-05", "bibliographicIssueDateType": "Issued"}, "bibliographicPageEnd": "833", "bibliographicPageStart": "823", "bibliographicVolumeNumber": "35", "bibliographic_titles": [{"bibliographic_title": "Annals of Nuclear Medicine"}]}]}, "item_8_description_5": {"attribute_name": "抄録", "attribute_value_mlt": [{"subitem_description": "Objective: To examine the impact of acquisition time on Lutetium-177 (177Lu) single-photon emission computed tomography (SPECT) images using Monte Carlo simulation.\n\nMethods: A gamma camera simulation based on the Monte Carlo method was performed to produce SPECT images. The phantom was modeled on a NEMA IEC BODY phantom including six spheres as tumors. After the administration of 7.4 GBq of 177Lu, radioactivity concentrations of the tumor/liver at 6, 24, and 72 h after administration were set to 1.85/0.201, 2.12/0.156, and 1.95/0.117 MBq/mL, respectively. In addition, the radioactivity concentrations of the tumor at 72 h after administration varied by 1/2, 1/4, and 1/8 when comparison was made. Acquisition times examined were 1.2, 1.5, 2, 3, 6, and 12 min. To assess the impact of collimators, SPECT data acquired at 72 h after the administration using six collimators of low-energy high-resolution (LEHR), extended low-energy general-purpose (ELEGP), medium-energy, and general-purpose (MEGP-1, MEGP-2, and MEGP-3) and high-energy general-purpose (HEGP) were examined. After prefiltering using a Butterworth filter, projection images were reconstructed using ordered subset expectation maximization. The detected photons were classified into direct rays, scattered rays, penetrating rays, and characteristic X-rays from lead. The image quality was evaluated through visual assessment, and physical assessment of contrast recovery coefficient (CRC) and contrast-to-noise ratio (CNR). In this study, the CNR threshold for detectability was assumed to be 5.0.\n\nResults: To compare collimators, the highest sensitivity was observed with ELEGP, followed by LEHR and MEGP-1. The highest ratio of direct ray was also observed in ELEGP followed by MEGP-1. In comparison of the radioactivity concentration ratios of tumor/liver, CRC and CNR were significantly decreased with smaller radioactivity concentration ratios. This effect was greater with larger spheres. According to the visual assessment, the acquisition time of 6, 6, and 3 min or longer was required using ELEGP collimator at 6, 24, and 72 h after administration, respectively. Physical assessment based on CNR and CRC also suggested that 6, 6, and 3 min or longer acquisition time was necessary at 6, 24, and 72 h after administration.\n\nConclusion: 177Lu-SPECT images generated via the Monte Carlo simulation suggested that the recommended acquisition time was 6 min or longer at 6 and 24 h and 3 min or longer at 72 h after administration.", "subitem_description_type": "Abstract"}]}, "item_8_publisher_8": {"attribute_name": "出版者", "attribute_value_mlt": [{"subitem_publisher": "Springer Nature"}]}, "item_8_relation_13": {"attribute_name": "PubMed番号", "attribute_value_mlt": [{"subitem_relation_type_id": {"subitem_relation_type_id_text": "34057655", "subitem_relation_type_select": "PMID"}}]}, "item_8_relation_14": {"attribute_name": "DOI", "attribute_value_mlt": [{"subitem_relation_type_id": {"subitem_relation_type_id_text": "10.1007/s12149-021-01620-9", "subitem_relation_type_select": "DOI"}}]}, "item_8_relation_17": {"attribute_name": "関連サイト", "attribute_value_mlt": [{"subitem_relation_type_id": {"subitem_relation_type_id_text": "https://link.springer.com/article/10.1007%2Fs12149-021-01620-9", "subitem_relation_type_select": "URI"}}]}, "item_8_source_id_9": {"attribute_name": "ISSN", "attribute_value_mlt": [{"subitem_source_identifier": "0914-7187", "subitem_source_identifier_type": "ISSN"}]}, "item_access_right": {"attribute_name": "アクセス権", "attribute_value_mlt": [{"subitem_access_right": "metadata only access", "subitem_access_right_uri": "http://purl.org/coar/access_right/c_14cb"}]}, "item_creator": {"attribute_name": "著者", "attribute_type": "creator", "attribute_value_mlt": [{"creatorNames": [{"creatorName": "Yuya, Sekikawa"}], "nameIdentifiers": [{"nameIdentifier": "997028", "nameIdentifierScheme": "WEKO"}]}, {"creatorNames": [{"creatorName": "Keita, Funada"}], "nameIdentifiers": [{"nameIdentifier": "997029", "nameIdentifierScheme": "WEKO"}]}, {"creatorNames": [{"creatorName": "Go, Akamatsu"}], "nameIdentifiers": [{"nameIdentifier": "997030", "nameIdentifierScheme": "WEKO"}]}, {"creatorNames": [{"creatorName": "Kazuhiko, Himuro"}], "nameIdentifiers": [{"nameIdentifier": "997031", "nameIdentifierScheme": "WEKO"}]}, {"creatorNames": [{"creatorName": "Akihiko, Takahashi"}], "nameIdentifiers": [{"nameIdentifier": "997032", "nameIdentifierScheme": "WEKO"}]}, {"creatorNames": [{"creatorName": "Shingo, Baba"}], "nameIdentifiers": [{"nameIdentifier": "997033", "nameIdentifierScheme": "WEKO"}]}, {"creatorNames": [{"creatorName": "Masayuki, Sasaki"}], "nameIdentifiers": [{"nameIdentifier": "997034", "nameIdentifierScheme": "WEKO"}]}, {"creatorNames": [{"creatorName": "Go, Akamatsu", "creatorNameLang": "en"}], "nameIdentifiers": [{"nameIdentifier": "997035", "nameIdentifierScheme": "WEKO"}]}]}, "item_language": {"attribute_name": "言語", "attribute_value_mlt": [{"subitem_language": "eng"}]}, "item_resource_type": {"attribute_name": "資源タイプ", "attribute_value_mlt": [{"resourcetype": "journal article", "resourceuri": "http://purl.org/coar/resource_type/c_6501"}]}, "item_title": "Monte Carlo simulation of the acquisition conditions for 177Lu molecular imaging of hepatic tumors.", "item_titles": {"attribute_name": "タイトル", "attribute_value_mlt": [{"subitem_title": "Monte Carlo simulation of the acquisition conditions for 177Lu molecular imaging of hepatic tumors."}]}, "item_type_id": "8", "owner": "1", "path": ["1"], "permalink_uri": "https://repo.qst.go.jp/records/82943", "pubdate": {"attribute_name": "公開日", "attribute_value": "2021-06-04"}, "publish_date": "2021-06-04", "publish_status": "0", "recid": "82943", "relation": {}, "relation_version_is_last": true, "title": ["Monte Carlo simulation of the acquisition conditions for 177Lu molecular imaging of hepatic tumors."], "weko_shared_id": -1}
Monte Carlo simulation of the acquisition conditions for 177Lu molecular imaging of hepatic tumors.
https://repo.qst.go.jp/records/82943
https://repo.qst.go.jp/records/829436a08b85f-3854-4682-9dda-82ccd62e6d12
Item type | 学術雑誌論文 / Journal Article(1) | |||||
---|---|---|---|---|---|---|
公開日 | 2021-06-04 | |||||
タイトル | ||||||
タイトル | Monte Carlo simulation of the acquisition conditions for 177Lu molecular imaging of hepatic tumors. | |||||
言語 | ||||||
言語 | eng | |||||
資源タイプ | ||||||
資源タイプ識別子 | http://purl.org/coar/resource_type/c_6501 | |||||
資源タイプ | journal article | |||||
アクセス権 | ||||||
アクセス権 | metadata only access | |||||
アクセス権URI | http://purl.org/coar/access_right/c_14cb | |||||
著者 |
Yuya, Sekikawa
× Yuya, Sekikawa× Keita, Funada× Go, Akamatsu× Kazuhiko, Himuro× Akihiko, Takahashi× Shingo, Baba× Masayuki, Sasaki× Go, Akamatsu |
|||||
抄録 | ||||||
内容記述タイプ | Abstract | |||||
内容記述 | Objective: To examine the impact of acquisition time on Lutetium-177 (177Lu) single-photon emission computed tomography (SPECT) images using Monte Carlo simulation. Methods: A gamma camera simulation based on the Monte Carlo method was performed to produce SPECT images. The phantom was modeled on a NEMA IEC BODY phantom including six spheres as tumors. After the administration of 7.4 GBq of 177Lu, radioactivity concentrations of the tumor/liver at 6, 24, and 72 h after administration were set to 1.85/0.201, 2.12/0.156, and 1.95/0.117 MBq/mL, respectively. In addition, the radioactivity concentrations of the tumor at 72 h after administration varied by 1/2, 1/4, and 1/8 when comparison was made. Acquisition times examined were 1.2, 1.5, 2, 3, 6, and 12 min. To assess the impact of collimators, SPECT data acquired at 72 h after the administration using six collimators of low-energy high-resolution (LEHR), extended low-energy general-purpose (ELEGP), medium-energy, and general-purpose (MEGP-1, MEGP-2, and MEGP-3) and high-energy general-purpose (HEGP) were examined. After prefiltering using a Butterworth filter, projection images were reconstructed using ordered subset expectation maximization. The detected photons were classified into direct rays, scattered rays, penetrating rays, and characteristic X-rays from lead. The image quality was evaluated through visual assessment, and physical assessment of contrast recovery coefficient (CRC) and contrast-to-noise ratio (CNR). In this study, the CNR threshold for detectability was assumed to be 5.0. Results: To compare collimators, the highest sensitivity was observed with ELEGP, followed by LEHR and MEGP-1. The highest ratio of direct ray was also observed in ELEGP followed by MEGP-1. In comparison of the radioactivity concentration ratios of tumor/liver, CRC and CNR were significantly decreased with smaller radioactivity concentration ratios. This effect was greater with larger spheres. According to the visual assessment, the acquisition time of 6, 6, and 3 min or longer was required using ELEGP collimator at 6, 24, and 72 h after administration, respectively. Physical assessment based on CNR and CRC also suggested that 6, 6, and 3 min or longer acquisition time was necessary at 6, 24, and 72 h after administration. Conclusion: 177Lu-SPECT images generated via the Monte Carlo simulation suggested that the recommended acquisition time was 6 min or longer at 6 and 24 h and 3 min or longer at 72 h after administration. |
|||||
書誌情報 |
Annals of Nuclear Medicine 巻 35, p. 823-833, 発行日 2021-05 |
|||||
出版者 | ||||||
出版者 | Springer Nature | |||||
ISSN | ||||||
収録物識別子タイプ | ISSN | |||||
収録物識別子 | 0914-7187 | |||||
PubMed番号 | ||||||
識別子タイプ | PMID | |||||
関連識別子 | 34057655 | |||||
DOI | ||||||
識別子タイプ | DOI | |||||
関連識別子 | 10.1007/s12149-021-01620-9 | |||||
関連サイト | ||||||
識別子タイプ | URI | |||||
関連識別子 | https://link.springer.com/article/10.1007%2Fs12149-021-01620-9 |