量研学術機関リポジトリ「QST-Repository」は、国立研究開発法人 量子科学技術研究開発機構に所属する職員等が生み出した学術成果(学会誌発表論文、学会発表、研究開発報告書、特許等)を集積しインターネット上で広く公開するサービスです。 Welcome to QST-Repository where we accumulates and discloses the academic research results(Journal Publications, Conference presentation, Research and Development Report, Patent, etc.) of the members of National Institutes for Quantum Science and Technology.
Thank you very much for using our website. On the 11th of March 2019, this site was moved from our own network server to the JAIRO Cloud network server. If you previously bookmarked this site, that bookmark will no longer work. We would be grateful if you could bookmark the website again. Thank you very much for your understanding and cooperation.
We have been developing a medical imaging technique using a Compton camera. This study evaluates the feasibility of clear imaging with 99mTc and 18F simultaneously, and demonstrates in vivo imaging with 99mTc and/or 18F. We used a Compton camera with silicon and cadmium telluride (Si/CdTe) semiconductors. We estimated the imaging performance of the Compton camera for 141 keV and 511 keV gamma rays from 99mTc and 22Na, respectively. Next, we simultaneously imaged 99mTc and 18F point sources to evaluate the cross-talk artifacts produced by a higher energy gamma-ray background. Then, in the in vivo experiments, three rats were injected with 99mTc-dimercaptosuccinic acid and/or 18F-fluorodeoxyglucose and imaged. The Compton images were compared with PET images. The rats were euthanized, and the activities in their organs were measured using a well counter. The energy resolution and spatial resolution were measured for the sources. No apparent cross-talk artifacts were observed in the practical-activity ratio (99mTc:18F = 1:16). We succeeded in imaging the distributions of 99mTc and 18F simultaneously, and the results were consistent with the PET images and well counter measurements. Our Si/CdTe Compton camera can thus work as a multi-tracer imager, covering various SPECT and PET probes, with less cross-talk artifacts in comparison to the conventional Anger cameras using a collimator. Our findings suggest the possibility of human trials.