Virus Detection using Second Harmonics of Magnetic Nanoparticles

https://repo.qst.go.jp/records/85565
https://repo.qst.go.jp/records/85565
dadf1e24-1616-4420-b4dc-ee4e418492e1
Item type 学術雑誌論文 / Journal Article(1)
公開日 2022-03-22
タイトル
タイトル Virus Detection using Second Harmonics of Magnetic Nanoparticles
言語
言語 eng
資源タイプ
資源タイプ識別子 http://purl.org/coar/resource_type/c_6501
資源タイプ journal article
アクセス権
アクセス権 metadata only access
アクセス権URI http://purl.org/coar/access_right/c_14cb
著者 Hirota, Ryuichi

× Hirota, Ryuichi

WEKO 1041450

Hirota, Ryuichi

Search repository
Murayama, Toru

× Murayama, Toru

WEKO 1041451

Murayama, Toru

Search repository
Katsumi, Ryota

× Katsumi, Ryota

WEKO 1041452

Katsumi, Ryota

Search repository
Kawawaki, Tokuhisa

× Kawawaki, Tokuhisa

WEKO 1041453

Kawawaki, Tokuhisa

Search repository
Yabukami, Shin

× Yabukami, Shin

WEKO 1041454

Yabukami, Shin

Search repository
Ryuji, Igarashi

× Ryuji, Igarashi

WEKO 1041455

Ryuji, Igarashi

Search repository
Negishi, Yuichi

× Negishi, Yuichi

WEKO 1041456

Negishi, Yuichi

Search repository
Kusakabe, Moriaki

× Kusakabe, Moriaki

WEKO 1041457

Kusakabe, Moriaki

Search repository
Sekino, Masaki

× Sekino, Masaki

WEKO 1041458

Sekino, Masaki

Search repository
Yatsui, Takashi

× Yatsui, Takashi

WEKO 1041459

Yatsui, Takashi

Search repository
Kuwahata, Akihiro

× Kuwahata, Akihiro

WEKO 1041460

Kuwahata, Akihiro

Search repository
Ryuji, Igarashi

× Ryuji, Igarashi

WEKO 1041461

en Ryuji, Igarashi

Search repository
抄録
内容記述タイプ Abstract
内容記述 Virus detection based on the nonlinear magnetic response of magnetic nanoparticles have been investigated, and magnetic detection methods using third harmonic have been widely applied because of high sensitivity, short measurement-time, and low cost. In this letter, we propose a virus detection method using the second harmonic signal to improve the signal intensity. Our results show that the signal to noise ratio, SNR, of the second harmonic is approximately three times higher than that of the third harmonic. A comparison of the ratio of the second harmonic to the fourth harmonic R24 and the ratio of the third harmonic to the fifth harmonic R35 shows that R24 is more sensitive in identifying changes in virus concentration. Our method has the potential to be used for rapid screening on virus detection.
書誌情報 IEEJ Transactions on Electrical and Electronic Engineering

巻 17, 号 8, 発行日 2022-04
出版者
出版者 Wiley Online Library
ISSN
収録物識別子タイプ ISSN
収録物識別子 1931-4973
DOI
識別子タイプ DOI
関連識別子 10.1002/tee.23613
関連サイト
識別子タイプ DOI
関連識別子 https://doi.org/10.1002/tee.23613
戻る
0
views
See details
Views

Versions

Ver.1 2023-05-15 16:54:06.133995
Show All versions

Share

Mendeley Twitter Facebook Print Addthis

Cite as

エクスポート

OAI-PMH
  • OAI-PMH JPCOAR 2.0
  • OAI-PMH JPCOAR 1.0
  • OAI-PMH DublinCore
  • OAI-PMH DDI
Other Formats
  • JSON
  • BIBTEX

Confirm


Powered by WEKO3


Powered by WEKO3