WEKO3
アイテム
Development of crosshair light sharing PET detector with TOF and DOI capabilities using fast LGSO scintillator
https://repo.qst.go.jp/records/85178
https://repo.qst.go.jp/records/85178aebcc8e1-7d43-407f-90b1-5de1d9538f44
Item type | 学術雑誌論文 / Journal Article(1) | |||||
---|---|---|---|---|---|---|
公開日 | 2021-12-28 | |||||
タイトル | ||||||
タイトル | Development of crosshair light sharing PET detector with TOF and DOI capabilities using fast LGSO scintillator | |||||
言語 | ||||||
言語 | eng | |||||
資源タイプ | ||||||
資源タイプ識別子 | http://purl.org/coar/resource_type/c_6501 | |||||
資源タイプ | journal article | |||||
アクセス権 | ||||||
アクセス権 | metadata only access | |||||
アクセス権URI | http://purl.org/coar/access_right/c_14cb | |||||
著者 |
Eiji, Yoshida
× Eiji, Yoshida× Fujino, Obata× Kamada, Kei× Yoshikawa, Akira× Taiga, Yamaya× Eiji, Yoshida× Fujino, Obata× Taiga, Yamaya |
|||||
抄録 | ||||||
内容記述タイプ | Abstract | |||||
内容記述 | Objective. Time-of-flight (TOF) and depth-of-interaction (DOI) are well recognized as important information to improve PET image quality. Since such information types are not correlated, many TOF-DOI detectors have been developed but there are only a few reports of high-resolution detectors (e.g. 1.5 mm resolution) for brain PET systems. Based on the DOI detector, which enables single-ended readout by optically coupling a pair of crystals and having a loop structure, we have developed the crosshair light sharing (CLS) PET detector that optically couples the four-loop structure, consisting of quadrisected crystals comparable in size to a photo-sensor, to four photo-sensors in close proximity arranged in a windmill shape. Even as a high-resolution detector, the CLS PET detector could obtain both TOF and DOI information. The coincidence resolving time (CRT) of the CLS PET detector needs to be further improved, however, for application to the brain PET system. Recently, a fast LGSO crystal was developed which has advantages in detection efficiency and CRT compared to the GFAG crystal. In this work, we developed the CLS PET detector using the fast LGSO crystal for the TOF-DOI brain PET system. Approach. The crystals were each 1.45 × 1.45 × 15 mm3 and all surfaces were chemically etched. The CLS PET detector consisted of a 14 × 14 crystal array optically coupled to an 8 × 8 MPPC array. Main results. The fast LGSO array provided 10.1% energy resolution at 511 keV, 4.7 mm DOI resolution at 662 keV, and 293 ps CRT with the energy window of 440–620 keV. Significance. The developed CLS PET detector has 290% higher coincidence sensitivity, 30% better energy resolution, and 32% better time resolution compared to our previous CLS PET detector. | |||||
書誌情報 |
Physics in Medicine & Biology 巻 66, 号 22, p. 225003, 発行日 2021-11 |
|||||
出版者 | ||||||
出版者 | IOP Publishing | |||||
ISSN | ||||||
収録物識別子タイプ | ISSN | |||||
収録物識別子 | 0031-9155 | |||||
DOI | ||||||
識別子タイプ | DOI | |||||
関連識別子 | 10.1088/1361-6560/ac2f8b | |||||
関連サイト | ||||||
識別子タイプ | URI | |||||
関連識別子 | https://iopscience.iop.org/article/10.1088/1361-6560/ac2f8b |