Development of crosshair light sharing PET detector with TOF and DOI capabilities using fast LGSO scintillator

https://repo.qst.go.jp/records/85178
https://repo.qst.go.jp/records/85178
aebcc8e1-7d43-407f-90b1-5de1d9538f44
Item type 学術雑誌論文 / Journal Article(1)
公開日 2021-12-28
タイトル
タイトル Development of crosshair light sharing PET detector with TOF and DOI capabilities using fast LGSO scintillator
言語
言語 eng
資源タイプ
資源タイプ識別子 http://purl.org/coar/resource_type/c_6501
資源タイプ journal article
アクセス権
アクセス権 metadata only access
アクセス権URI http://purl.org/coar/access_right/c_14cb
著者 Eiji, Yoshida

× Eiji, Yoshida

WEKO 1026753

Eiji, Yoshida

Search repository
Fujino, Obata

× Fujino, Obata

WEKO 1026754

Fujino, Obata

Search repository
Kamada, Kei

× Kamada, Kei

WEKO 1026755

Kamada, Kei

Search repository
Yoshikawa, Akira

× Yoshikawa, Akira

WEKO 1026756

Yoshikawa, Akira

Search repository
Taiga, Yamaya

× Taiga, Yamaya

WEKO 1026757

Taiga, Yamaya

Search repository
Eiji, Yoshida

× Eiji, Yoshida

WEKO 1026758

en Eiji, Yoshida

Search repository
Fujino, Obata

× Fujino, Obata

WEKO 1026759

en Fujino, Obata

Search repository
Taiga, Yamaya

× Taiga, Yamaya

WEKO 1026760

en Taiga, Yamaya

Search repository
抄録
内容記述タイプ Abstract
内容記述 Objective. Time-of-flight (TOF) and depth-of-interaction (DOI) are well recognized as important information to improve PET image quality. Since such information types are not correlated, many TOF-DOI detectors have been developed but there are only a few reports of high-resolution detectors (e.g. 1.5 mm resolution) for brain PET systems. Based on the DOI detector, which enables single-ended readout by optically coupling a pair of crystals and having a loop structure, we have developed the crosshair light sharing (CLS) PET detector that optically couples the four-loop structure, consisting of quadrisected crystals comparable in size to a photo-sensor, to four photo-sensors in close proximity arranged in a windmill shape. Even as a high-resolution detector, the CLS PET detector could obtain both TOF and DOI information. The coincidence resolving time (CRT) of the CLS PET detector needs to be further improved, however, for application to the brain PET system. Recently, a fast LGSO crystal was developed which has advantages in detection efficiency and CRT compared to the GFAG crystal. In this work, we developed the CLS PET detector using the fast LGSO crystal for the TOF-DOI brain PET system. Approach. The crystals were each 1.45 × 1.45 × 15 mm3 and all surfaces were chemically etched. The CLS PET detector consisted of a 14 × 14 crystal array optically coupled to an 8 × 8 MPPC array. Main results. The fast LGSO array provided 10.1% energy resolution at 511 keV, 4.7 mm DOI resolution at 662 keV, and 293 ps CRT with the energy window of 440–620 keV. Significance. The developed CLS PET detector has 290% higher coincidence sensitivity, 30% better energy resolution, and 32% better time resolution compared to our previous CLS PET detector.
書誌情報 Physics in Medicine & Biology

巻 66, 号 22, p. 225003, 発行日 2021-11
出版者
出版者 IOP Publishing
ISSN
収録物識別子タイプ ISSN
収録物識別子 0031-9155
DOI
識別子タイプ DOI
関連識別子 10.1088/1361-6560/ac2f8b
関連サイト
識別子タイプ URI
関連識別子 https://iopscience.iop.org/article/10.1088/1361-6560/ac2f8b
戻る
0
views
See details
Views

Versions

Ver.1 2023-05-15 17:03:20.032213
Show All versions

Share

Mendeley Twitter Facebook Print Addthis

Cite as

エクスポート

OAI-PMH
  • OAI-PMH JPCOAR 2.0
  • OAI-PMH JPCOAR 1.0
  • OAI-PMH DublinCore
  • OAI-PMH DDI
Other Formats
  • JSON
  • BIBTEX

Confirm


Powered by WEKO3


Powered by WEKO3