Strategy to Reduce the Collective Equivalent Dose for the Lens of the Physician’s Eye Using Short Radiation Protection Curtains to Prevent Cataracts

https://repo.qst.go.jp/records/84425
https://repo.qst.go.jp/records/84425
2d1fd103-cab9-426c-ac4c-71f996a47bfb
Item type 学術雑誌論文 / Journal Article(1)
公開日 2021-10-14
タイトル
タイトル Strategy to Reduce the Collective Equivalent Dose for the Lens of the Physician’s Eye Using Short Radiation Protection Curtains to Prevent Cataracts
言語
言語 eng
資源タイプ
資源タイプ識別子 http://purl.org/coar/resource_type/c_6501
資源タイプ journal article
アクセス権
アクセス権 metadata only access
アクセス権URI http://purl.org/coar/access_right/c_14cb
著者 Nakagami, Koichi

× Nakagami, Koichi

WEKO 1019312

Nakagami, Koichi

Search repository
Moritake, Takashi

× Moritake, Takashi

WEKO 1019313

Moritake, Takashi

Search repository
Nagamoto, Keisuke

× Nagamoto, Keisuke

WEKO 1019314

Nagamoto, Keisuke

Search repository
Morota, Koichi

× Morota, Koichi

WEKO 1019315

Morota, Koichi

Search repository
Matsuzaki, Satoru

× Matsuzaki, Satoru

WEKO 1019316

Matsuzaki, Satoru

Search repository
Kuriyama, Tomoko

× Kuriyama, Tomoko

WEKO 1019317

Kuriyama, Tomoko

Search repository
Kunugita, Naoki

× Kunugita, Naoki

WEKO 1019318

Kunugita, Naoki

Search repository
Takashi, Moritake

× Takashi, Moritake

WEKO 1019319

en Takashi, Moritake

Search repository
Naoki, Kunugita

× Naoki, Kunugita

WEKO 1019320

en Naoki, Kunugita

Search repository
抄録
内容記述タイプ Abstract
内容記述 A short curtain that improves on the low versatility of existing long curtains was developed as a dedicated radiation protective device for the over-table tube fluorographic imaging units. The effect of this short curtain in preventing cataracts was then examined. First, the physician lens dose reduction rate was obtained at the position of the lens. Next, the reduction rate in the collective equivalent dose for the lens of the physician's eye was estimated. The results showed that lens dose reduction rates with the long curtain and the short curtain were 88.9% (literature-based value) and 17.6%, respectively, higher with the long curtain. In our hospital, the reduction rate in the collective equivalent dose for the lens of the physician's eye was 9.8% and 17.6% with a procedures mixture, using the long curtain where technically possible and no curtain in all other procedures, and the short curtain in all procedures, respectively, higher with the short curtain. Moreover, a best available for curtains raised the reduction rate in the collective equivalent dose for the lens of the physician's eye a maximum of 25.5%. By introducing the short curtain, it can be expected to have an effect in preventing cataracts in medical staff.
書誌情報 Diagnostics

巻 11, 号 8, p. 1415, 発行日 2021-08
出版者
出版者 MDPI
ISSN
収録物識別子タイプ ISSN
収録物識別子 2075-4418
PubMed番号
識別子タイプ PMID
関連識別子 34441349
DOI
識別子タイプ DOI
関連識別子 10.3390/diagnostics11081415
関連サイト
識別子タイプ URI
関連識別子 https://www.mdpi.com/2075-4418/11/8/1415
戻る
0
views
See details
Views

Versions

Ver.1 2023-05-15 17:11:12.342261
Show All versions

Share

Mendeley Twitter Facebook Print Addthis

Cite as

エクスポート

OAI-PMH
  • OAI-PMH JPCOAR 2.0
  • OAI-PMH JPCOAR 1.0
  • OAI-PMH DublinCore
  • OAI-PMH DDI
Other Formats
  • JSON
  • BIBTEX

Confirm


Powered by WEKO3


Powered by WEKO3