Heavy-ion Radiotherapy Rotating Gantry Calibration for Improving Patient Positioning Accuracy

https://repo.qst.go.jp/records/84068
https://repo.qst.go.jp/records/84068
8a0bbbd9-c0df-4e48-901b-4f6072a497ef
Item type 会議発表用資料 / Presentation(1)
公開日 2021-11-12
タイトル
タイトル Heavy-ion Radiotherapy Rotating Gantry Calibration for Improving Patient Positioning Accuracy
言語
言語 eng
資源タイプ
資源タイプ識別子 http://purl.org/coar/resource_type/c_c94f
資源タイプ conference object
アクセス権
アクセス権 metadata only access
アクセス権URI http://purl.org/coar/access_right/c_14cb
著者 Yukinobu, Sakata

× Yukinobu, Sakata

WEKO 1015343

Yukinobu, Sakata

Search repository
Ryusuke, Hirai

× Ryusuke, Hirai

WEKO 1015344

Ryusuke, Hirai

Search repository
Shinichiro, Mori

× Shinichiro, Mori

WEKO 1015345

Shinichiro, Mori

Search repository
Yukinobu, Sakata

× Yukinobu, Sakata

WEKO 1015346

en Yukinobu, Sakata

Search repository
Ryusuke, Hirai

× Ryusuke, Hirai

WEKO 1015347

en Ryusuke, Hirai

Search repository
Shinichiro, Mori

× Shinichiro, Mori

WEKO 1015348

en Shinichiro, Mori

Search repository
抄録
内容記述タイプ Abstract
内容記述 Purpose/Objective(s)

In heavy-ion radiotherapy, accurate patient positioning is necessary for treatment accuracy. For this purpose, calibration of the X-ray imaging device position is indispensable. However, because positional repeatability of a calibration phantom in the rotating gantry affects the calibration result, it is not possible to calibrate using known 3-D positions of iron balls. Hence, instead of using the positions of the iron balls, we developed an X-ray tube position calibration method using the position and rotation of the FPD measured by a laser tracker. In the proposed method, the positions of the X-ray tube and iron balls are adjusted so that the distance between the position to be projected of each iron ball and the detected position (reprojection error) is minimized.

Materials/Methods

In the rotating gantry calibration, multiple images are taken while changing the gantry angle, but since the position of the calibration phantom during imaging is unchanged, it is possible to represent the 3-D iron ball position with common parameters among the images. As described above, bundle adjustment can be cited as a method for calculating the position of the 3-D point and the imaging system parameter from the corresponding point between the multiple view images. The principle is to adjust all the unknown parameters so that the projected iron ball position on the image by the estimated parameters matches as closely as possible the iron ball position detected from the image. In this paper, we propose a rotating gantry calibration method based on bundle adjustment. First, by approximating the position of the X-ray tube at which the reprojection error becomes small as a function of the iron ball position of the calibration phantom, the number of parameters is reduced and the position of the iron ball is optimized. Next, the positions of the iron balls are fixed and the position of the X-ray tube is optimized. At this time, by analytically calculating the solution of two axes out of three axes where reprojection error becomes the minimum, efficient search is executed as a 1-D optimization problem. A head phantom was imaged while rotating the gantry in increments of 15 degrees, patient positioning was performed at each angle, and how the result changes with or without calibration was evaluated.


Results

For the head phantom study, the Euclidean distance patient positioning error of each angle with respect to the positioning result at the reference angle was 0.62 ± 0.21 mm (mean ± SD) and the maximum was about 1.0 mm without calibration, but with calibration, these values became 0.09 ± 0.03 mm and 0.2 mm.

Conclusion

We proposed an automatic calibration method for rotating gantry heavy-ion radiotherapy. By using the method, the patient positioning accuracy of the rotating gantry was improved.
会議概要(会議名, 開催地, 会期, 主催者等)
内容記述タイプ Other
内容記述 The Amercian Society for Radiation Oncology (ASTRO) Annual meeting 2019
発表年月日
日付 2019-09-16
日付タイプ Issued
戻る
0
views
See details
Views

Versions

Ver.1 2023-05-15 17:15:58.445567
Show All versions

Share

Mendeley Twitter Facebook Print Addthis

Cite as

エクスポート

OAI-PMH
  • OAI-PMH JPCOAR 2.0
  • OAI-PMH JPCOAR 1.0
  • OAI-PMH DublinCore
  • OAI-PMH DDI
Other Formats
  • JSON
  • BIBTEX

Confirm


Powered by WEKO3


Powered by WEKO3