Experimental Assessment on Data Sufficiency Condition for WGI Compton Imaging
https://repo.qst.go.jp/records/83759
https://repo.qst.go.jp/records/837591ae71e9a-90e5-4edb-84c5-30127626e0e8
Item type | 会議発表用資料 / Presentation(1) | |||||
---|---|---|---|---|---|---|
公開日 | 2021-10-06 | |||||
タイトル | ||||||
タイトル | Experimental Assessment on Data Sufficiency Condition for WGI Compton Imaging | |||||
言語 | ||||||
言語 | eng | |||||
資源タイプ | ||||||
資源タイプ識別子 | http://purl.org/coar/resource_type/c_c94f | |||||
資源タイプ | conference object | |||||
アクセス権 | ||||||
アクセス権 | metadata only access | |||||
アクセス権URI | http://purl.org/coar/access_right/c_14cb | |||||
著者 |
Hideaki, Tashima
× Hideaki, Tashima× Eiji, Yoshida× Sodai, Takyu× Fumihiko, Nishikido× Takumi, Nishina× Mikio, Suga× Hidekatsu, Wakizaka× Miwako, Takahashi× Kotaro, Nagatsu× Atsushi, Tsuji× Kamada, Kei× Yoshikawa, Akira× Parodi, Katia× Taiga, Yamaya× Hideaki, Tashima× Eiji, Yoshida× Sodai, Takyu× Fumihiko, Nishikido× Takumi, Nishina× Mikio, Suga× Hidekatsu, Wakizaka× Miwako, Takahashi× Kotaro, Nagatsu× Atsushi, Tsuji× Parodi, Katia× Taiga, Yamaya |
|||||
抄録 | ||||||
内容記述タイプ | Abstract | |||||
内容記述 | We developed a whole gamma imaging (WGI) prototype composed of a scatterer ring insert and an absorber ring. It was the world’s first realization of a full-ring Compton camera geometry. In this study, we focused on the data sufficiency condition (DSC) for Compton image reconstruction. Contrary to PET which requires a full-ring geometry to satisfy the DSC, Compton imaging does not require a full ring theoretically. Reducing the detector-ring coverages can improve flexibility of the geometry design and reduce manufacturing costs although sensitivity is decreased. However, the DSC for reconstruction is not guaranteed in a practical situation due to actual absorber ring arrangement and detection limits in scattering angle. In this study, we experimentally assessed the DSC for the angular coverages. We evaluated image artifacts on reconstructed Compton images because, in general, images distort and exhibit artifacts when the DSC is not satisfied, degrading quantitative accuracy. We measured a cylindrical phantom filled with 89Zr solution of 10.2 MBq by the WGI prototype for 60 min and extracted 909-keV Compton events. We selected events related to the detector blocks marked as available to change detector angular coverages defined for the phantom bottom point. First, we changed the number of detectors in the scatterer ring, keeping the full-ring absorber. Next, we changed the absorber ring angular coverage by changing the number of detectors, keeping the scatterer ring at the number where the angular coverage just exceeded 180°. As a result, when the angular coverage of the scatterer detector was less than 180°, reconstructed images exhibited artifacts. The absorber ring detector with less than 180° coverage showed artifacts when used with the scatterer detector with more than 180° coverage. We concluded that angular coverages of more than 180° for both the scatterer and absorber detectors against any region on a subject are the DSC for Compton image reconstruction. | |||||
会議概要(会議名, 開催地, 会期, 主催者等) | ||||||
内容記述タイプ | Other | |||||
内容記述 | IEEE NSS MIC 2021 | |||||
発表年月日 | ||||||
日付 | 2021-10-20 | |||||
日付タイプ | Issued |