Development and characterization of optical readout well-type glass gas electron multiplier for dose imaging in clinical carbon beams
https://repo.qst.go.jp/records/81830
https://repo.qst.go.jp/records/81830cc75d47f-92f5-4ba2-aa9c-d3f5eeb9e342
Item type | 学術雑誌論文 / Journal Article(1) | |||||
---|---|---|---|---|---|---|
公開日 | 2021-02-15 | |||||
タイトル | ||||||
タイトル | Development and characterization of optical readout well-type glass gas electron multiplier for dose imaging in clinical carbon beams | |||||
言語 | ||||||
言語 | eng | |||||
資源タイプ | ||||||
資源タイプ識別子 | http://purl.org/coar/resource_type/c_6501 | |||||
資源タイプ | journal article | |||||
アクセス権 | ||||||
アクセス権 | metadata only access | |||||
アクセス権URI | http://purl.org/coar/access_right/c_14cb | |||||
著者 |
Fujiwara, Takeshi
× Fujiwara, Takeshi× Yusuke, Koba× Mitsuya, Yuki× Nakamura, Riichiro× Tatsumoto, Ryuta× Kawahara, Shuto× Maehata, Keisuke× Yamaguchi, Hidetoshi× Chang, Weishan× Naruhiro, Matsufuji× Takahashi, Hiroyuki× Yusuke, Koba× Chang, Weishan× Naruhiro, Matsufuji |
|||||
抄録 | ||||||
内容記述タイプ | Abstract | |||||
内容記述 | The use of carbon ion beams in cancer therapy (also known as hadron therapy) is steadily growing worldwide; therefore, the demand for more efficient dosimetry systems is also increasing because daily quality assurance (QA) measurements of hadron radiotherapy is one of the most complex and time consuming tasks. The aim of this study is to develop a two-dimensional dosimetry system that offers high spatial resolution, a large field of view, quick data response, and a linear dose–response relationship. We demonstrate the dose imaging performance of a novel digital dose imager using carbon ion beams for hadron therapy. The dose imager is based on a newly-developed gaseous detector, a well-type glass gas electron multiplier. The imager is successfully operated in a hadron therapy facility with clinical intensity beams for radiotherapy. It features a high spatial resolution of less than 1 mm and an almost linear dose–response relationship with no saturation and very low linear-energy-transfer dependence. Experimental results show that the dose imager has the potential to improve dosimetry accuracy for daily QA. |
|||||
書誌情報 |
Physica Medica 巻 82, p. 72-78, 発行日 2021-02 |
|||||