タイトル | Highly sensitive detection of sodium in aqueous solutions using laser-induced breakdown spectroscopy with liquid sheet jets | |||||
言語 | ||||||
言語 | eng | |||||
資源タイプ | ||||||
資源タイプ識別子 | http://purl.org/coar/resource_type/c_6501 | |||||
資源タイプ | journal article | |||||
アクセス権 | ||||||
アクセス権 | metadata only access | |||||
アクセス権URI | http://purl.org/coar/access_right/c_14cb | |||||
著者 |
Nakanishi, Ryuzo
× Nakanishi, Ryuzo× Oba, Hironori× Saeki, Morihisa× Wakaida, Ikuo× Rie, Tanabe-Yamagishi× Ito, Yoshiro× Ryuzo, Nakanishi× Hironori, Oba× Morihisa, Saeki |
|||||
抄録 | ||||||
内容記述タイプ | Abstract | |||||
内容記述 | Laser-induced breakdown spectroscopy (LIBS) combined with liquid jets was applied to the detection of trace sodium (Na) in aqueous solutions. The sensitivities of two types of liquid jets were compared: a liquid cylindrical jet with a diameter of 500 μm and a liquid sheet jet with a thickness of 20 μm. Compared with the cylindrical jet, the liquid sheet jet effectively reduced the splash from the laser-irradiated surface and produced long-lived luminous plasma. The limit of detection (LOD) of Na was determined to be 0.57 µg/L for the sheet jet and 10.5 µg/L for the cylindrical jet. The LOD obtained for the sheet jet was comparable to those obtained for commercially available inductively coupled plasma emission spectrometers. | |||||
書誌情報 |
Optics Express 巻 29, 号 4, p. 5205-5212, 発行日 2021-02 |
|||||
出版者 | ||||||
出版者 | The Optical Society | |||||
ISSN | ||||||
収録物識別子タイプ | ISSN | |||||
収録物識別子 | 1094-4087 | |||||
DOI | ||||||
識別子タイプ | DOI | |||||
関連識別子 | 10.1364/OE.415308 | |||||
関連サイト | ||||||
識別子タイプ | DOI | |||||
関連識別子 | https://doi.org/10.1364/OE.415308 |