ログイン
言語:

WEKO3

  • トップ
  • ランキング
To
lat lon distance
To

Field does not validate



インデックスリンク

インデックスツリー

メールアドレスを入力してください。

WEKO

One fine body…

WEKO

One fine body…

アイテム

  1. プロシーディングス

Rapid Pu analysis by ICP-MS and alpha spectrometry in urine samples for patient screening after radiation accident

https://repo.qst.go.jp/records/80179
https://repo.qst.go.jp/records/80179
4614f16d-3988-4a86-a5b8-bb21285fcbd7
Item type 会議発表論文 / Conference Paper(1)
公開日 2020-07-15
タイトル
タイトル Rapid Pu analysis by ICP-MS and alpha spectrometry in urine samples for patient screening after radiation accident
言語
言語 eng
資源タイプ
資源タイプ識別子 http://purl.org/coar/resource_type/c_5794
資源タイプ conference paper
アクセス権
アクセス権 metadata only access
アクセス権URI http://purl.org/coar/access_right/c_14cb
著者 Yang, Guosheng

× Yang, Guosheng

WEKO 996299

Yang, Guosheng

Search repository
Zheng, Jian

× Zheng, Jian

WEKO 996300

Zheng, Jian

Search repository
Eunjoo, Kim

× Eunjoo, Kim

WEKO 996301

Eunjoo, Kim

Search repository
Seno, Hatsuho

× Seno, Hatsuho

WEKO 996302

Seno, Hatsuho

Search repository
Kurihara, Osamu

× Kurihara, Osamu

WEKO 996303

Kurihara, Osamu

Search repository
Yang, Guosheng

× Yang, Guosheng

WEKO 996304

en Yang, Guosheng

Search repository
Zheng, Jian

× Zheng, Jian

WEKO 996305

en Zheng, Jian

Search repository
Eunjoo, Kim

× Eunjoo, Kim

WEKO 996306

en Eunjoo, Kim

Search repository
Hatsuho, Seno

× Hatsuho, Seno

WEKO 996307

en Hatsuho, Seno

Search repository
Osamu, Kurihara

× Osamu, Kurihara

WEKO 996308

en Osamu, Kurihara

Search repository
抄録
内容記述タイプ Abstract
内容記述 For the 2019 PROCORAD (Association for the PROmotion of Quality COntrol in RADiotoxicological Analysis) intercomparison campaign for urine actinides analysis, we compared SF-ICP-MS and ICP-MS/MS combined with chromatographic separation using TEVA+UTEVA+DGA resins and alpha spectrometry combined with chromatographic separation using TEVA resin for the measurement of 239Pu. The Pu yields were in the range of 68–93 % (n=20). Both the methods of SF-ICP-MS (2.44±0.05 mBq/sample for 19 ACTA, n=4 and 7.63±0.13 mBq/sample for 19 ACTB, n=4) and ICP-MS/MS (2.45±0.21 mBq/sample for 19 ACTA, n=1 and 7.89±0.54 mBq/sample for 19 ACTB, n=1) presented precise and accurate results that agreed well with the assigned values (2.52±0.26 mBq/sample for 19 ACTA and 7.56±0.39 mBq/sample for 19 ACTB). The method of alpha spectrometry, showing larger uncertainties, also presented comparable results (3.49±1.29 mBq/sample for 19 ACTA, n=1 and 8.05±1.76 mBq/sample for 19 ACTB, n=1). Due to the high sensitivity of SF-ICP-MS and ICP-MS/MS, rapid determination of 239Pu (ca. 10 h) can be achieved in small volume urine sample (20 mL) generally. In contrast, the method of alpha spectrometry generally consumes more sample (e.g. 500 mL) and is more time-consuming. Therefore, it would be more feasible to measure long-lived 239Pu and 240Pu by the methods of SF-ICP-MS and ICP-MS/MS, and short-lived 238Pu by the method of alpha spectrometry for emergency response.
書誌情報 Proceedings of the 21th Workshop on Environmental Radioactivity (KEK proceedings)

p. 163-169, 発行日 2020-11
出版者
出版者 High Energy accelerator Research Organization (KEK)
戻る
0
views
See details
Views

Versions

Ver.1 2023-05-15 17:36:01.130197
Show All versions

Share

Mendeley Twitter Facebook Print Addthis

Cite as

エクスポート

OAI-PMH
  • OAI-PMH JPCOAR 2.0
  • OAI-PMH JPCOAR 1.0
  • OAI-PMH DublinCore
  • OAI-PMH DDI
Other Formats
  • JSON
  • BIBTEX

Confirm


Powered by WEKO3