ログイン
言語:

WEKO3

  • トップ
  • ランキング
To
lat lon distance
To

Field does not validate



インデックスリンク

インデックスツリー

メールアドレスを入力してください。

WEKO

One fine body…

WEKO

One fine body…

アイテム

  1. 学会発表・講演等
  2. ポスター発表

Production of carbon ion at electron cyclotron resonance ion source for high energy carbon ion radiotherapy

https://repo.qst.go.jp/records/78241
https://repo.qst.go.jp/records/78241
20e7b1f1-ad14-42d0-952e-b85b58251cb4
Item type 会議発表用資料 / Presentation(1)
公開日 2019-12-26
タイトル
タイトル Production of carbon ion at electron cyclotron resonance ion source for high energy carbon ion radiotherapy
言語
言語 eng
資源タイプ
資源タイプ識別子 http://purl.org/coar/resource_type/c_c94f
資源タイプ conference object
アクセス権
アクセス権 metadata only access
アクセス権URI http://purl.org/coar/access_right/c_14cb
著者 Muramatsu, Masayuki

× Muramatsu, Masayuki

WEKO 824782

Muramatsu, Masayuki

Search repository
Muramatsu, Masayuki

× Muramatsu, Masayuki

WEKO 824783

en Muramatsu, Masayuki

Search repository
抄録
内容記述タイプ Abstract
内容記述 Carbon-ion radiotherapy (C-ion RT) using Heavy Ion Medical Accelerator in Chiba (HIMAC) has been carried out in National Institute of Radiological Sciences (NIRS-QST) since 1994. Over 12000 cancer patients have been treated with carbon beams having energies of between 56-430 MeV/ u. A 10 GHz electron cyclotron resonance ion source, named NIRS-ECR, is used for production of carbon ion at HIMAC. There is a carbon stacking in the plasma chamber and insulator for beam extraction. Therefore, we needed cleaning of NIRS-ECR two times per year until September 2018. In order to increase an operation time for C-ion RT, we change operation parameters for production carbon ion. Usually, CH4 gas is used for production of carbon ion. We decreased the gas flow with enough beam current for medical use. As a result, we could operate during one year with small stacking of carbon to the insulator compared previous parameters. However, we found new problem about starting of the NIRS-ECR. Plasma is not turn on during about 10 minutes in starting. We increase the gas flow a little since September 2019. Improvement of carbon ion production with long life time at NIRS-ECR for C-ion RT will be described.
会議概要(会議名, 開催地, 会期, 主催者等)
内容記述タイプ Other
内容記述 Accelerator Reliability Workshop 2019 (ARW2019)
発表年月日
日付 2019-11-12
日付タイプ Issued
戻る
0
views
See details
Views

Versions

Ver.1 2023-05-15 18:28:24.189908
Show All versions

Share

Mendeley Twitter Facebook Print Addthis

Cite as

エクスポート

OAI-PMH
  • OAI-PMH JPCOAR 2.0
  • OAI-PMH JPCOAR 1.0
  • OAI-PMH DublinCore
  • OAI-PMH DDI
Other Formats
  • JSON
  • BIBTEX

Confirm


Powered by WEKO3


Powered by WEKO3