Thermal Neutron Flux Measurement by Counting Conversion Electrons from 134mCs Generated in a CsI Scintillator
https://repo.qst.go.jp/records/77784
https://repo.qst.go.jp/records/777846be1fc3c-8161-417d-92a9-ff78840ef127
Item type | 会議発表論文 / Conference Paper(1) | |||||
---|---|---|---|---|---|---|
公開日 | 2019-12-03 | |||||
タイトル | ||||||
タイトル | Thermal Neutron Flux Measurement by Counting Conversion Electrons from 134mCs Generated in a CsI Scintillator | |||||
言語 | ||||||
言語 | eng | |||||
資源タイプ | ||||||
資源タイプ識別子 | http://purl.org/coar/resource_type/c_5794 | |||||
資源タイプ | conference paper | |||||
アクセス権 | ||||||
アクセス権 | metadata only access | |||||
アクセス権URI | http://purl.org/coar/access_right/c_14cb | |||||
著者 |
Shimazu, Miyu
× Shimazu, Miyu× Genichiro Wakabayashi× Akihiro Nohtomi× Yusuke Koba× Kiyomitsu Shinsho× Koba, Yusuke |
|||||
抄録 | ||||||
内容記述タイプ | Abstract | |||||
内容記述 | A new method to measure low intensity neutrons such as secondary neutrons generated in radiotherapy has been developed. In the proposed method, thermal neutron flux is determined by measuring the conversion electrons from 134m Cs generated in a CsI(Tl) scintillator irradiated by neutrons. Because the generated 134m Cs is contained in the scintillator as internal sources, the emitted conversion electrons fully stop in the scintillator and counted with high efficiency. The thermal neutron flux in the neutron radiography facility of the Kindai University's research reactor, UTR-KINKI, was measured by the proposed method. The cadmium difference method was used to separate the contributions of thermal and epi-thermal neutrons. The result was in good agreement with the one obtained by the conventional gold foil activation method. It was concluded that the proposed method is useful for the measurement of thermal neutron flux. | |||||
書誌情報 |
2018 IEEE Nuclear Science Symposium and Medical Imaging Conference Proceedings p. 1-3, 発行日 2019-09 |
|||||
DOI | ||||||
識別子タイプ | DOI | |||||
関連識別子 | 10.1109/NSSMIC.2018.8824515 |