WEKO3
アイテム
Accurate and precise determination of 90Sr at femtogram level in IAEA proficiency test using Thermal Ionization Mass Spectrometry
https://repo.qst.go.jp/records/77463
https://repo.qst.go.jp/records/774630f047f99-a1d4-4be6-a77c-1c6c2ab891d1
Item type | 学術雑誌論文 / Journal Article(1) | |||||
---|---|---|---|---|---|---|
公開日 | 2019-11-18 | |||||
タイトル | ||||||
タイトル | Accurate and precise determination of 90Sr at femtogram level in IAEA proficiency test using Thermal Ionization Mass Spectrometry | |||||
言語 | ||||||
言語 | eng | |||||
資源タイプ | ||||||
資源タイプ識別子 | http://purl.org/coar/resource_type/c_6501 | |||||
資源タイプ | journal article | |||||
アクセス権 | ||||||
アクセス権 | metadata only access | |||||
アクセス権URI | http://purl.org/coar/access_right/c_14cb | |||||
著者 |
Kavasi, Norbert
× Kavasi, Norbert× Kumar Sahoo, Sarata× Arae, Hideki× Aono, Tatsuo× Palacz, Zenon× Kavasi, Norbert× Sahoo, Sarata× Hideki, Arae× Tatsuo, Aono |
|||||
抄録 | ||||||
内容記述タイプ | Abstract | |||||
内容記述 | A novel method for the determination of ultra-trace level 90Sr has been recently developed applying thermal ionization mass spectrometry (TIMS). The method includes the chemical separation of Zr (isobaric interference of 90Zr) from the samples followed by determination of 90Sr/88Sr abundance sensitivity (2.1 × 10−10). The analytical performance of this method was assessed in the IAEA-TEL 2017-3 worldwide open proficiency test. For 90Sr determination, tap water and milk powder samples were distributed amongst the participant laboratories with reference values of 11.2 ± 0.3 Bq kg−1 (2.2 ± 0.1 fg g−1) and 99.9 ± 5.0 Bq kg−1 (19.5 ± 1.0 fg g−1), respectively. The stable Sr concentrations were 39.4 ± 0.9 ng g−1 and 2.5 ± 0.1 μg g−1 while the 90Sr/88Sr isotope ratios were 6.47 ± 0.17 × 10−8 and 9.04 ± 0.45 × 10−9 in the tap water and milk powder samples, respectively. For TIMS measurement, 50 mL water and 1 g milk powder samples were taken for analysis. This TIMS method demonstrated an impressive accuracy (relative bias of 4.2% and −2.1%, respectively) and precision (relative combined uncertainty of 4.1% and 7.6%, respectively) when compared with radiometric techniques. For the first time in the history of inorganic mass-spectrometry, 90Sr analysis using a TIMS instrument is confirmed by an independent proficiency test. |
|||||
書誌情報 |
Scientific Reports 巻 9, p. 16532-1, 発行日 2019-11 |
|||||
出版者 | ||||||
出版者 | Nature research | |||||
ISSN | ||||||
収録物識別子タイプ | ISSN | |||||
収録物識別子 | 2045-2322 | |||||
DOI | ||||||
識別子タイプ | DOI | |||||
関連識別子 | 10.1038/s41598-019-52890-3 | |||||
関連サイト | ||||||
識別子タイプ | URI | |||||
関連識別子 | https://www.nature.com/articles/s41598-019-52890-3 |