ログイン
言語:

WEKO3

  • トップ
  • ランキング
To
lat lon distance
To

Field does not validate



インデックスリンク

インデックスツリー

メールアドレスを入力してください。

WEKO

One fine body…

WEKO

One fine body…

アイテム

  1. 原著論文

A walk-in type calibration chamber facility for 222Rn measuring devices and inter-comparison exercises

https://repo.qst.go.jp/records/77283
https://repo.qst.go.jp/records/77283
7c998f7a-52cf-4ec7-afa0-66c32767e8ae
Item type 学術雑誌論文 / Journal Article(1)
公開日 2019-11-01
タイトル
タイトル A walk-in type calibration chamber facility for 222Rn measuring devices and inter-comparison exercises
言語
言語 eng
資源タイプ
資源タイプ識別子 http://purl.org/coar/resource_type/c_6501
資源タイプ journal article
アクセス権
アクセス権 metadata only access
アクセス権URI http://purl.org/coar/access_right/c_14cb
著者 Trilochana, S

× Trilochana, S

WEKO 868518

Trilochana, S

Search repository
M Somashekarappa, H

× M Somashekarappa, H

WEKO 868519

M Somashekarappa, H

Search repository
Sudeep Kumara, K

× Sudeep Kumara, K

WEKO 868520

Sudeep Kumara, K

Search repository
P Mohan, M

× P Mohan, M

WEKO 868521

P Mohan, M

Search repository
Rashmi Nayak, S

× Rashmi Nayak, S

WEKO 868522

Rashmi Nayak, S

Search repository
Renita, Shiny D’Souza

× Renita, Shiny D’Souza

WEKO 868523

Renita, Shiny D’Souza

Search repository
S Kamath, Srinivas

× S Kamath, Srinivas

WEKO 868524

S Kamath, Srinivas

Search repository
K Sahoo, B

× K Sahoo, B

WEKO 868525

K Sahoo, B

Search repository
J Gaware, J

× J Gaware, J

WEKO 868526

J Gaware, J

Search repository
K Sapra, B

× K Sapra, B

WEKO 868527

K Sapra, B

Search repository
Janik, Miroslaw

× Janik, Miroslaw

WEKO 868528

Janik, Miroslaw

Search repository
Darwish, Al-Azmi

× Darwish, Al-Azmi

WEKO 868529

Darwish, Al-Azmi

Search repository
S Mayya, Y

× S Mayya, Y

WEKO 868530

S Mayya, Y

Search repository
Karunakara, N

× Karunakara, N

WEKO 868531

Karunakara, N

Search repository
Janik, Miroslaw

× Janik, Miroslaw

WEKO 868532

en Janik, Miroslaw

Search repository
抄録
内容記述タイプ Abstract
内容記述 A walk-in type 222Rn calibration chamber of volume 22.7 m3, which has traceability to international standards, is established at the Centre for Advanced Research in Environmental Radioactivity, Mangalore University, India. It has a human–machine interface communication system, a programmable logic controller and sensor feedback circuit for controlling and data acquisition of relative humidity (RH) and temperature (T). An innovative method for the generation of desired 222Rn concentration (a few hundred Bq m−3 up to about 36 kBq m−3) using soil gas as a source was adopted. Leak rates of 222Rn from the chamber for the mixing fan ON and OFF conditions were determined to be 0.0011 and 0.00018 h−1 respectively. With the exhaust system fully turned on, the maximum clearance rate of the chamber was 0.58 ± 0.07 h−1. Excellent spatial uniformity in 222Rn concentration in the chamber was confirmed (with a mean value of relative standard deviation < 12%) through measurements at 23 locations using CR-39 film-based passive devices. Demonstration of calibration applications was performed using charcoal canister and PicoRad vials as the 222Rn adsorption devices. The study shows that gamma spectrometry is a convenient alternative approach to liquid scintillation analysis of PicoRad vials for 222Rn measurement.
書誌情報 Radiation Protection Dosimetry

巻 187, 号 4, p. 466-481, 発行日 2019-10
出版者
出版者 Oxford University Press
ISSN
収録物識別子タイプ ISSN
収録物識別子 0144-8420
DOI
識別子タイプ DOI
関連識別子 10.1093/rpd/ncz188
関連サイト
識別子タイプ URI
関連識別子 https://academic.oup.com/rpd/advance-article/doi/10.1093/rpd/ncz188/5608969
戻る
0
views
See details
Views

Versions

Ver.1 2023-05-15 18:07:33.263545
Show All versions

Share

Mendeley Twitter Facebook Print Addthis

Cite as

エクスポート

OAI-PMH
  • OAI-PMH JPCOAR 2.0
  • OAI-PMH JPCOAR 1.0
  • OAI-PMH DublinCore
  • OAI-PMH DDI
Other Formats
  • JSON
  • BIBTEX

Confirm


Powered by WEKO3