Optimization of a Staggered 3-Layer DOI PET Detector with a 1 mm LYSO Pitch for High-Resolution Small Animal PET Imaging

https://repo.qst.go.jp/records/77148
https://repo.qst.go.jp/records/77148
0a8ea597-fd12-4b6c-99bf-1388f54ed022
Item type 会議発表用資料 / Presentation(1)
公開日 2019-10-16
タイトル
タイトル Optimization of a Staggered 3-Layer DOI PET Detector with a 1 mm LYSO Pitch for High-Resolution Small Animal PET Imaging
言語
言語 eng
資源タイプ
資源タイプ識別子 http://purl.org/coar/resource_type/c_c94f
資源タイプ conference object
アクセス権
アクセス権 metadata only access
アクセス権URI http://purl.org/coar/access_right/c_14cb
著者 Kang, Hangyu

× Kang, Hangyu

WEKO 793876

Kang, Hangyu

Search repository
Nishikido, Fumihiko

× Nishikido, Fumihiko

WEKO 793877

Nishikido, Fumihiko

Search repository
Yamaya, Taiga

× Yamaya, Taiga

WEKO 793878

Yamaya, Taiga

Search repository
Kang, Hangyu

× Kang, Hangyu

WEKO 793879

en Kang, Hangyu

Search repository
Nishikido, Fumihiko

× Nishikido, Fumihiko

WEKO 793880

en Nishikido, Fumihiko

Search repository
Yamaya, Taiga

× Yamaya, Taiga

WEKO 793881

en Yamaya, Taiga

Search repository
抄録
内容記述タイプ Abstract
内容記述 Small animal positron emission tomography (PET) requires a submillimeter resolution for better quantification of radiopharmaceuticals. The aim of this study was to develop a staggered 3-layer depth-of-interaction (DOI) PET detector with a 1 mm crystal pitch for realizing a 0.5 mm small animal PET scanner of 52 mm ring diameter. The proposed small animal DOI PET detector had 3-layer staggered LYSO crystal arrays, an acrylic light guide with a 1 mm thickness, and a 4×4 SiPM (Hamamatsu, S13361-3050NE-04, Japan) array with a pixel pitch of 3.2 mm. The LYSO crystal had dimensions of 0.9 × 0.9 × 5.0 mm3 with a pitch of 1 mm. The 1st (10 × 9), 2nd (10 × 10), and 3rd (11 × 11) LYSO layers were stacked with a staggered configuration to encode DOI information in the 2D crystal map. In order to investigate the optimal reflector material and light guide thickness, the performance of the PET detector module was evaluated with different reflectors (ESR and BaSO4), and different light guide thicknesses (0.5, 1.0, and 2.0 mm). Sixteen SiPM anode signals were multiplexed into four positional signals by using a resistive network. The SiPM analog signals were digitized by CAMAC DAQ. The BaSO4 reflector resulted in better crystal identification than the ESR did. However, the energy resolution with the BaSO4 reflector (15.8±2.0%) was worse than that of the ESR reflector (11.1±1.0%). The acrylic light guide thickness of 1 mm resulted in the best crystal identification for both BaSO4 and ESR reflectors. In conclusion, a staggered 3-layer DOI PET detector with a 1 mm crystal pitch was developed by using a 1 mm thick light guide and BaSO4 reflector. In the near future, a prototype high-resolution small animal DOI PET scanner will be developed.
会議概要(会議名, 開催地, 会期, 主催者等)
内容記述タイプ Other
内容記述 2019 IEEE Nuclear Science Symposium and Medical Imaging Conference
発表年月日
日付 2019-10-31
日付タイプ Issued
戻る
0
views
See details
Views

Versions

Ver.1 2023-05-15 18:40:41.217183
Show All versions

Share

Mendeley Twitter Facebook Print Addthis

Cite as

エクスポート

OAI-PMH
  • OAI-PMH JPCOAR 2.0
  • OAI-PMH JPCOAR 1.0
  • OAI-PMH DublinCore
  • OAI-PMH DDI
Other Formats
  • JSON
  • BIBTEX

Confirm


Powered by WEKO3


Powered by WEKO3