A staggered three-layer DOI PET detector with a 1 mm crystal pitch for a high resolution small animal PET imaging

https://repo.qst.go.jp/records/75999
https://repo.qst.go.jp/records/75999
3b4b4efe-50ed-4ec2-84e5-e8354feca98f
Item type 会議発表用資料 / Presentation(1)
公開日 2019-06-07
タイトル
タイトル A staggered three-layer DOI PET detector with a 1 mm crystal pitch for a high resolution small animal PET imaging
言語
言語 eng
資源タイプ
資源タイプ識別子 http://purl.org/coar/resource_type/c_c94f
資源タイプ conference object
アクセス権
アクセス権 metadata only access
アクセス権URI http://purl.org/coar/access_right/c_14cb
著者 Kang, Hangyu

× Kang, Hangyu

WEKO 760736

Kang, Hangyu

Search repository
Takyu, Sodai

× Takyu, Sodai

WEKO 760737

Takyu, Sodai

Search repository
Nishikido, Fumihiko

× Nishikido, Fumihiko

WEKO 760738

Nishikido, Fumihiko

Search repository
Yamaya, Taiga

× Yamaya, Taiga

WEKO 760739

Yamaya, Taiga

Search repository
Kang, Hangyu

× Kang, Hangyu

WEKO 760740

en Kang, Hangyu

Search repository
Takyu, Sodai

× Takyu, Sodai

WEKO 760741

en Takyu, Sodai

Search repository
Nishikido, Fumihiko

× Nishikido, Fumihiko

WEKO 760742

en Nishikido, Fumihiko

Search repository
Yamaya, Taiga

× Yamaya, Taiga

WEKO 760743

en Yamaya, Taiga

Search repository
抄録
内容記述タイプ Abstract
内容記述 Objectives: Development of a staggered three-layer DOI PET detector with a 1 mm crystal pitch for realizing a 0.7 mm resolution small animal PET imager of 52 mm ring diameter.
Methods: The proposed small animal DOI PET detector consists of three-layer staggered LYSO crystal arrays, an acrylic light guide with a thickness of 1 mm, and 4×4 SiPM (Hamamatsu, S13360-3050NE-04, Japan) array with a pixel pitch of 3.2 mm. The LYSO crystal had a dimension of 0.9 × 0.9 × 5.0 mm3 with a pitch of 1 mm. The 1st (10 × 9), the 2nd (10 × 10), and the 3rd (11 × 11) LYSO layers were stacked with a staggered configuration to provide DOI information in the 2D crystal map. In order to investigate the optimal reflector and light guide thickness, the performance of the PET detector module was evaluated with different reflectors (ESR and BaSO4), and different light guide thicknesses (0.5, 1.0, and 2.0 mm). A 22Na source was used to irradiate 511 keV photon to the detector module. The sixteen SiPM output signals were multiplexed into four position signals by using the Anger logic circuit, and a trigger signal was generated by using the common cathode signals. The SiPM signals were digitized by CAMAC DAQ.
Results: The BaSO4 reflector resulted in better crystal identification for the three-layer staggered LYSO crystal arrays as compared to the ESR. However, the energy resolution with BaSO4 (38.8%) was worse than that of the ESR (20.3%). The acrylic light guide thickness of 1 mm resulted in the best crystal identification for both BaSO4 and ESR reflector cases.
Conclusions: A staggered three-layer DOI PET detector with a 1 mm crystal pitch was developed. All the LYSO crystals could be identified clearly on the 2D flood histogram by using the BaSO4 reflector. In future, a prototype high resolution small animal DOI PET scanner will be developed by using the proposed three-layer staggered DOI PET detector module.
会議概要(会議名, 開催地, 会期, 主催者等)
内容記述タイプ Other
内容記述 2019 Annual Meeting of Society of Nuclear Medicine and Moldecular Imaging
発表年月日
日付 2019-06-25
日付タイプ Issued
戻る
0
views
See details
Views

Versions

Ver.1 2023-05-15 18:53:10.711366
Show All versions

Share

Mendeley Twitter Facebook Print Addthis

Cite as

エクスポート

OAI-PMH
  • OAI-PMH JPCOAR 2.0
  • OAI-PMH JPCOAR 1.0
  • OAI-PMH DublinCore
  • OAI-PMH DDI
Other Formats
  • JSON
  • BIBTEX

Confirm


Powered by WEKO3


Powered by WEKO3