WEKO3
アイテム
Derivation and Validity Evaluation of Calibration Factors for Activated-charcoal Radon Collectors
https://repo.qst.go.jp/records/75831
https://repo.qst.go.jp/records/758312ec18dc5-72ef-4d5c-93d4-29a1def9677f
Item type | 学術雑誌論文 / Journal Article(1) | |||||
---|---|---|---|---|---|---|
公開日 | 2019-05-17 | |||||
タイトル | ||||||
タイトル | Derivation and Validity Evaluation of Calibration Factors for Activated-charcoal Radon Collectors | |||||
言語 | ||||||
言語 | eng | |||||
資源タイプ | ||||||
資源タイプ識別子 | http://purl.org/coar/resource_type/c_6501 | |||||
資源タイプ | journal article | |||||
アクセス権 | ||||||
アクセス権 | metadata only access | |||||
アクセス権URI | http://purl.org/coar/access_right/c_14cb | |||||
著者 |
Wakabayashi, Akiko
× Wakabayashi, Akiko× Yasuoka, Yumi× Janik, Miroslaw× Nagahama, Hiroyuki× Fukuhori, Nobutoshi× Mori, Yasunori× Arai, Yuria× Fujii, Satomi× Mukai, Takahiro× Janik, Miroslaw |
|||||
抄録 | ||||||
内容記述タイプ | Abstract | |||||
内容記述 | Radon collectors (e.g., PicoRad collectors) based on activated-charcoal have been used for screening and measuring radon. However, researchers at the U. S. Environmental Protection Agency reported that they could not verify the proper functioning of some commercially available radon detectors, including PicoRad collectors. In this study, we exposed two lots (with different expiration dates) of PicoRad collectors to the reference conditions at a controlled radon concentration within a radon chamber because the calibration factors were derived for use by a semi-empirical equation. Further, we exposed the PicoRad collectors to an uncontrolled radon atmosphere for conducting validity evaluation. The radon concentration results obtained by the Pico-Rad collectors using the semi-empirical equation were observed to be in good agreement with the conventional true radon concentration value. It denoted the optimal value of a quantity determined using a reference instrument. These experiments revealed that different values were required for the factors of the conversion equation of each radon collector lot with a different expiration date. | |||||
書誌情報 |
RADIOISOTOPES 巻 68, p. 317-329, 発行日 2019-05 |
|||||
出版者 | ||||||
出版者 | Japan Radioisotope Association | |||||
ISSN | ||||||
収録物識別子タイプ | ISSN | |||||
収録物識別子 | 0033-8303 | |||||
DOI | ||||||
識別子タイプ | DOI | |||||
関連識別子 | 10.3769/radioisotopes.68.317 | |||||
関連サイト | ||||||
識別子タイプ | URI | |||||
関連識別子 | https://academic.oup.com/rpd/advance-article/doi/10.1093/rpd/ncz100/5490187 |