  1. 原著論文

PRELIMINARY TECHNICAL DISCUSSION ON A NEW RADON AND ITS PROGENY CONTINUOUS MONITOR USING TWO-FILTER METHOD

https://repo.qst.go.jp/records/75755
https://repo.qst.go.jp/records/75755
7d1d244c-0ec1-4b50-a197-af70d023e501
Item type 学術雑誌論文 / Journal Article(1)
公開日 2019-05-07
タイトル
タイトル PRELIMINARY TECHNICAL DISCUSSION ON A NEW RADON AND ITS PROGENY CONTINUOUS MONITOR USING TWO-FILTER METHOD
言語
言語 eng
資源タイプ
資源タイプ識別子 http://purl.org/coar/resource_type/c_6501
資源タイプ journal article
アクセス権
アクセス権 metadata only access
アクセス権URI http://purl.org/coar/access_right/c_14cb
著者 五十嵐, 悠

× 五十嵐, 悠

WEKO 998462

五十嵐, 悠

Search repository
Nozaki , Takahiro

× Nozaki , Takahiro

WEKO 998463

Nozaki , Takahiro

Search repository
Mizuno , Hiroyuki

× Mizuno , Hiroyuki

WEKO 998464

Mizuno , Hiroyuki

Search repository
Kuroki , Tomohiro

× Kuroki , Tomohiro

WEKO 998465

Kuroki , Tomohiro

Search repository
内田, 祐棋

× 内田, 祐棋

WEKO 998466

内田, 祐棋

Search repository
ミロソラフ, ヤニック

× ミロソラフ, ヤニック

WEKO 998467

ミロソラフ, ヤニック

Search repository
Iimoto, Takeshi

× Iimoto, Takeshi

WEKO 998468

Iimoto, Takeshi

Search repository
Yu, Igarashi

× Yu, Igarashi

WEKO 998469

en Yu, Igarashi

Search repository
Yuuki, Uchida

× Yuuki, Uchida

WEKO 998470

en Yuuki, Uchida

Search repository
Janik, Miroslaw

× Janik, Miroslaw

WEKO 998471

en Janik, Miroslaw

Search repository
抄録
内容記述タイプ Abstract
内容記述 Focusing on the scalability of Two-Filter Method, we started to develop a monitor for the concentration of radon and its progeny. In this study, we investigated the influence of a high-humidity environment on measuring radon concentration, and the influence of the decay chamber of the monitor on the measurement. In the high-humidity test, the conversion factor of (Bqm−3)/(cpm) tends to raise with increasing humidity. On the measurement of radon progeny, existence of the decay chamber of the monitor makes measurement sensitivity lower under environments of little aerosols. Radon concentration measurement by the developed monitor could be influenced by environmental humidity, and that counting loss could occur due to deposition of radon progeny inside of the decay chamber. Correction relating these would be needed based on the data of calibration tests.
書誌情報 Radiation Protection Dosimetry

巻 184, 号 3-4, p. 418-421, 発行日 2019-05
出版者
出版者 Oxford University Press
ISSN
収録物識別子タイプ ISSN
収録物識別子 0144-8420
DOI
識別子タイプ DOI
関連識別子 10.1093/rpd/ncz072
関連サイト
識別子タイプ DOI
関連識別子 https://doi.org/10.1093/rpd/ncz072
戻る
0
views
See details
Views

Versions

Ver.1 2023-05-15 17:33:38.521220
Show All versions

Share

Mendeley Twitter Facebook Print Addthis

Cite as

エクスポート

OAI-PMH
  • OAI-PMH JPCOAR 2.0
  • OAI-PMH JPCOAR 1.0
  • OAI-PMH DublinCore
  • OAI-PMH DDI
Other Formats
  • JSON
  • BIBTEX

Confirm


Powered by WEKO3


Powered by WEKO3