WEKO3
アイテム
Scanned carbon-ion beam therapy throughput over the first 7 years at National Institute of Radiological Sciences
https://repo.qst.go.jp/records/73422
https://repo.qst.go.jp/records/73422351dc2c8-bec1-4f2c-a800-f3ea1d0a56d5
Item type | 学術雑誌論文 / Journal Article(1) | |||||
---|---|---|---|---|---|---|
公開日 | 2019-02-18 | |||||
タイトル | ||||||
タイトル | Scanned carbon-ion beam therapy throughput over the first 7 years at National Institute of Radiological Sciences | |||||
言語 | ||||||
言語 | eng | |||||
資源タイプ | ||||||
資源タイプ識別子 | http://purl.org/coar/resource_type/c_6501 | |||||
資源タイプ | journal article | |||||
アクセス権 | ||||||
アクセス権 | metadata only access | |||||
アクセス権URI | http://purl.org/coar/access_right/c_14cb | |||||
著者 |
森, 慎一郎
× 森, 慎一郎× 武井, 由佳× 白井, 敏之× 原, 洋介× 古川, 卓司× 稲庭, 拓× 谷本, 克之× 田尻, 稔× 黒岩, 大悟× 木村, 卓× 山本, 直敬× 山田, 滋× 辻, 比呂志× 鎌田, 正× Mori, Shinichiro× Takei, Yuka× Shirai, Toshiyuki× Hara, Yousuke× Furukawa, Takuji× Inaniwa, Taku× Tanimoto, Katsuyuki× Tajiri, Minoru× Kuroiwa, Daigo× Kimura, Taku× Yamamoto, Naoyoshi× Yamada, Shigeru× Tsuji, Hiroshi× Kamada, Tadashi |
|||||
抄録 | ||||||
内容記述タイプ | Abstract | |||||
内容記述 | Introduction: In the 7 years since our facility opened, we have treated > 2000 patients with pencil-beam scanned carbon-ion beam therapy. Methods: To summarize treatment workflow, we evaluated the following five metrics: i) total number of treated patients; ii) treatment planning time, not including contouring procedure; iii) quality assurance (QA) time (daily and patient-specific); iv) treatment room occupancy time, including patient setup, preparation time, and beam irradiation time; and v) daily treatment hours. These were derived from the oncology information system and patient handling system log files. Results: The annual number of treated patients reached 594, 7 years from the facility startup, using two treat- ment rooms. Mean treatment planning time was 6.0 h (minimum: 3.4 h for prostate, maximum: 9.3 h for eso- phagus). Mean time devoted to daily QA and patient-specific QA were 22 min and 13.5 min per port, respec- tively, for the irradiation beam system. Room occupancy time was 14.5 min without gating for the first year, improving to 9.2 min (8.2 min without gating and 12.8 min with gating) in the second. At full capacity, the system ran for 7.5 h per day. Conclusions: We are now capable of treating approximately 600 patients per year in two treatment rooms. Accounting for the staff working time, this performance appears reasonable compared to the other facilities. |
|||||
書誌情報 |
Physica Medica 巻 52, p. 18-26, 発行日 2018-06 |
|||||
出版者 | ||||||
出版者 | Elsevier | |||||
ISSN | ||||||
収録物識別子タイプ | ISSN | |||||
収録物識別子 | 1120-1797 | |||||
DOI | ||||||
識別子タイプ | DOI | |||||
関連識別子 | 10.1016/j.ejmp.2018.06.002 | |||||
関連サイト | ||||||
識別子タイプ | URI | |||||
関連識別子 | https://www.sciencedirect.com/science/article/abs/pii/S1120179718304836 |