Comparison of Sample Tubes for the X-band EPR Measurement of an aqueous sample: Effects on Reproducibility of Signal Intensities

https://repo.qst.go.jp/records/73031
https://repo.qst.go.jp/records/73031
d108a756-5607-4570-98db-543595739182
Item type 会議発表用資料 / Presentation(1)
公開日 2018-11-20
タイトル
タイトル Comparison of Sample Tubes for the X-band EPR Measurement of an aqueous sample: Effects on Reproducibility of Signal Intensities
言語
言語 eng
資源タイプ
資源タイプ識別子 http://purl.org/coar/resource_type/c_c94f
資源タイプ conference object
アクセス権
アクセス権 metadata only access
アクセス権URI http://purl.org/coar/access_right/c_14cb
著者 Makino, Yusuke

× Makino, Yusuke

WEKO 719696

Makino, Yusuke

Search repository
Fukui, Koji

× Fukui, Koji

WEKO 719697

Fukui, Koji

Search repository
Nakanishi, Ikuo

× Nakanishi, Ikuo

WEKO 719698

Nakanishi, Ikuo

Search repository
C., Krishna Murali

× C., Krishna Murali

WEKO 719699

C., Krishna Murali

Search repository
Matsumoto, Ken-ichiro

× Matsumoto, Ken-ichiro

WEKO 719700

Matsumoto, Ken-ichiro

Search repository
牧野 祐典

× 牧野 祐典

WEKO 719701

en 牧野 祐典

Search repository
中西 郁夫

× 中西 郁夫

WEKO 719702

en 中西 郁夫

Search repository
松本 謙一郎

× 松本 謙一郎

WEKO 719703

en 松本 謙一郎

Search repository
抄録
内容記述タイプ Abstract
内容記述 Reproducibility of the X-band EPR measurement of an aqueous solution sample was compared using three different types of sample tubes, and accuracy of quantitative performance was assessed.
A PTFE tubing, a glass capillary and a quartz flat cuvette were compared to get a suitable condition for quantitative measurement. An accurate 0.1 mM water solution of TEMPOL was used as a standard sample. The TEMPOL solution was loaded into one of sample tube, and the sample tube was set in the TE-mode cavity of X-band EPR spectrometer. Two procedures below were tested for reproducibility of repeated measurements. 1) The sample tube in the cavity was washed or renewed every measurement, and measurements were repeated several times with fixed EPR parameters. 2) The sample tubes in the cavity were stayed and repeatedly measured several times. Next, EPR parameters such as sweep speed, time constant, modulation width, and/or microwave power were varied to seek optimum signal intensity.
The PTFE tubing showed the best reproducibility when the measurements were repeated with staying the sample tube (the ratio of standard deviation to the averaged signal intensity; SD/AV = 0.0058). When the sample and sample tube was renewed every measurement, variation of signal intensity became larger (SD/AV = 0.0333). The glass capillary had the best reproducibility in both procedures 1 and 2 (SD/AV = 0.0104 and 0.0036, respectively). The signal reproducibility of the flat cuvette was relatively low (SD/AV = 0.0400 and 0.0690 for procedures 1 and 2, respectively). However, the flat cuvette gave the largest signal intensity when the loaded volume of the sample was the identical.
In conclusion, the best quantitative performance of the X-band EPR spectroscopy for a liquid sample was obtained when the measurements are carried out with the capillary.
会議概要(会議名, 開催地, 会期, 主催者等)
内容記述タイプ Other
内容記述 25th Annual Conference of the Society for Redox Biology and Medicine (SfRBM 2018)
発表年月日
日付 2018-11-15
日付タイプ Issued
戻る
0
views
See details
Views

Versions

Ver.1 2023-05-15 19:33:31.953977
Show All versions

Share

Mendeley Twitter Facebook Print Addthis

Cite as

エクスポート

OAI-PMH
  • OAI-PMH JPCOAR 2.0
  • OAI-PMH JPCOAR 1.0
  • OAI-PMH DublinCore
  • OAI-PMH DDI
Other Formats
  • JSON
  • BIBTEX

Confirm


Powered by WEKO3


Powered by WEKO3