Energy response correction for EPID dosimetry in photon radiation therapy

https://repo.qst.go.jp/records/72652
https://repo.qst.go.jp/records/72652
c8d86ed4-5d82-454b-baa3-3485e3c5452d
Item type 会議発表用資料 / Presentation(1)
公開日 2018-02-06
タイトル
タイトル Energy response correction for EPID dosimetry in photon radiation therapy
言語
言語 eng
資源タイプ
資源タイプ識別子 http://purl.org/coar/resource_type/c_c94f
資源タイプ conference object
アクセス権
アクセス権 metadata only access
アクセス権URI http://purl.org/coar/access_right/c_14cb
著者 Takei, Hideyuki

× Takei, Hideyuki

WEKO 715573

Takei, Hideyuki

Search repository
武居 秀行

× 武居 秀行

WEKO 715574

en 武居 秀行

Search repository
抄録
内容記述タイプ Abstract
内容記述 Purpose
Electronic portal imaging device (EPID) is a popular tool for a verification of the patient setup for radiation therapy. Recently, it is grabbing attention as a tool for in vivo dosimetry. It detects therapeutic photons which penetrated the patient body to acquire an image. Conversion to absorbed dose is required from the digital pixel value. However, the response of the EPID is susceptible to the energy spectrum changes through the body. The purpose of this study was to perform the energy response correction of the EPID image using Monte Carlo simulation (MC) to improve the accuracy of the EPID dosimetry.
Methods
Geant4.9.6 patch04 was used for the MC with a combination of a HTCondor 7.6 parallel computing software. The MC simulated the 6 and 10 MV photon beams from Trilogy (Varian Medical Systems) linear accelerator. The EPID response to the energy spectrum was evaluated. CT DICOM data set of a chest phantom was imported to the MC, and the simulated EPID images were acquired after irradiating 6 and 10 MV photon beams. The measurement was performed with the same setup as the MC and obtained the EPID images. The energy response correction was performed to the measured images to obtain the absorbed dose and compared with that measured with 2D array detector (PTW) as a reference.
Results
EPID was more sensitive to the photons penetrated the mediastinum than those penetrated the lung due to the contamination of the scattered photons with the energy of <0.1 MeV. The maximum differences between the doses measured with EPID and those measured with 2D array was 5.8% in mediastinum area with the 6 MV photon beam. The difference was less than 1.5% in average in lung area. The agreement of the corrected EPID dose and 2D array improved within 2.5% in mediastinum area for the 6 MV beam.
Conclusion
The EPID response to the energy spectrum was evaluated using MC. The corrected EPID dose agreed well with that measured with 2D array detector. The response correction using MC improved the accuracy of the EPID dosimetry.
会議概要(会議名, 開催地, 会期, 主催者等)
内容記述タイプ Other
内容記述 International conference on Monte Carlo techniques for medical applications (MCMA2017)
発表年月日
日付 2017-10-16
日付タイプ Issued
戻る
0
views
See details
Views

Versions

Ver.1 2023-05-15 19:37:30.968121
Show All versions

Share

Mendeley Twitter Facebook Print Addthis

Cite as

エクスポート

OAI-PMH
  • OAI-PMH JPCOAR 2.0
  • OAI-PMH JPCOAR 1.0
  • OAI-PMH DublinCore
  • OAI-PMH DDI
Other Formats
  • JSON
  • BIBTEX

Confirm


Powered by WEKO3


Powered by WEKO3