Influence of dose uncertainty with fractionation effect on tumor control probability in carbon-ion therapy

https://repo.qst.go.jp/records/71941
https://repo.qst.go.jp/records/71941
923197f2-a43c-412e-ab73-6cf2d51255ce
Item type 会議発表用資料 / Presentation(1)
公開日 2016-06-15
タイトル
タイトル Influence of dose uncertainty with fractionation effect on tumor control probability in carbon-ion therapy
言語
言語 eng
資源タイプ
資源タイプ識別子 http://purl.org/coar/resource_type/c_c94f
資源タイプ conference object
アクセス権
アクセス権 metadata only access
アクセス権URI http://purl.org/coar/access_right/c_14cb
著者 Sakama, Makoto

× Sakama, Makoto

WEKO 708584

Sakama, Makoto

Search repository
Kanematsu, Nobuyuki

× Kanematsu, Nobuyuki

WEKO 708585

Kanematsu, Nobuyuki

Search repository
Inaniwa, Taku

× Inaniwa, Taku

WEKO 708586

Inaniwa, Taku

Search repository
坂間 誠

× 坂間 誠

WEKO 708587

en 坂間 誠

Search repository
兼松 伸幸

× 兼松 伸幸

WEKO 708588

en 兼松 伸幸

Search repository
稲庭 拓

× 稲庭 拓

WEKO 708589

en 稲庭 拓

Search repository
抄録
内容記述タイプ Abstract
内容記述 In light ion radiotherapy, the dose concentration is highly sensitive to some uncertainties such as setup and range errors. Here we propose a method for evaluating the influence of dose uncertainty from spatial and range uncertainties with fractionation effect on tumor control probability (TCP) in carbon-ion therapy using the concept of equivalent stochastic dose, which was defined as a voxel dose that gives the mean expected survival fraction (SF) for the stochastically variable dose. Assuming that the probability density functions of dose and alpha variation for given errors are Gaussian distribution, the mean expected survival fractions were calculated with consideration of the correlation of setup and range errors between the fractions. Using this SF, equivalent stochastic clinical dose (ESCD) distributions, i.e. RBE-weighted dose distributions, of lung and prostate cases for various errors were calculated. To account for spatial dose heterogeneity, equivalent uniform stochastic clinical doses(EUSCD)were calculated by using the mean expected survival fraction in the volume of interest. Tumor control probability curves were calculated for assumed each error considering inter-patient sensitivity variation and dose uncertainty with fractionation effect, and affected by the contribution of setup and range errors and correlation of the errors. EUSCDs agreed well with TCP results except for case of large errors. The proposed evaluation method with EUSCD and TCP calculations will be useful for the stable indicator of tumor dose to improve treatment planning in carbon-ion therapy.
会議概要(会議名, 開催地, 会期, 主催者等)
内容記述タイプ Other
内容記述 The 55th Annual Conference of the Particle Therapy Co-operative Group (PTCOG55)
発表年月日
日付 2016-05-27
日付タイプ Issued
戻る
0
views
See details
Views