Development of 3D measurement device dedicated for range-compensator QA 3

https://repo.qst.go.jp/records/71754
https://repo.qst.go.jp/records/71754
db951383-557c-4069-89f0-3c9d021fe709
Item type 会議発表用資料 / Presentation(1)
公開日 2015-06-30
タイトル
タイトル Development of 3D measurement device dedicated for range-compensator QA 3
言語
言語 eng
資源タイプ
資源タイプ識別子 http://purl.org/coar/resource_type/c_c94f
資源タイプ conference object
アクセス権
アクセス権 metadata only access
アクセス権URI http://purl.org/coar/access_right/c_14cb
著者 Fukuda, Shigekazu

× Fukuda, Shigekazu

WEKO 706268

Fukuda, Shigekazu

Search repository
Shintani, Eriko

× Shintani, Eriko

WEKO 706269

Shintani, Eriko

Search repository
Yamada, Kiyohiro

× Yamada, Kiyohiro

WEKO 706270

Yamada, Kiyohiro

Search repository
Minohara, Shinichi

× Minohara, Shinichi

WEKO 706271

Minohara, Shinichi

Search repository
福田 茂一

× 福田 茂一

WEKO 706272

en 福田 茂一

Search repository
新谷 恵理子

× 新谷 恵理子

WEKO 706273

en 新谷 恵理子

Search repository
抄録
内容記述タイプ Abstract
内容記述 (Purpose)
In carbon therapy, the broad-beam method is used to irradiate beams to the tumor conformally. The broaden beams are cut to the shape of the target tumor projected in the beam’s eye view by a patient collimator or MLC. To adjust the range of the beams to the target in the depth a range-compensator is used. The range-compensator is a block that has an engraved depression in the shape of the target tumor. Though it is desirable to check the 3D shape of the range-compensator compared to the planned shape at all points, over 2000 points, in terms of QA/QC, only the limited points, 30-40 points, are usually checked by the contact length measurement sensor device (CLMSD). To verify all points automatically in a short time, the 3D measurement device (3DMD) dedicated for range-compensators QA has been developed.
(Methods)
The 3DMD has a XY-stage with high precision and a range-compensator is attached to the XY-stage. Two CCD laser displacement sensors were employed to measure the height of each point of the range-compensator by detecting reflected laser. The one sensor was changed to another type of sensor to improve the precision of the measurements. We verified some range-compensators.
(Results)
Due to the shape there were some points which cannot be measured, but the ratio of the number of the unmeasured points to that of all points was less than 1%. The precision of the measurements by the new sensor was improved and the concave positions were verified as acceptable.
(Conclusion)
After further commissioning of the 3DMD, we plan to use this device to check the range-compensators routinely.
会議概要(会議名, 開催地, 会期, 主催者等)
内容記述タイプ Other
内容記述 The 7th Korea-Japan Joint Meeting on Medical Physics (KJMP2014)
発表年月日
日付 2014-09-25
日付タイプ Issued
戻る
0
views
See details
Views

Versions

Ver.1 2023-05-15 19:47:46.385364
Show All versions

Share

Mendeley Twitter Facebook Print Addthis

Cite as

エクスポート

OAI-PMH
  • OAI-PMH JPCOAR 2.0
  • OAI-PMH JPCOAR 1.0
  • OAI-PMH DublinCore
  • OAI-PMH DDI
Other Formats
  • JSON
  • BIBTEX

Confirm


Powered by WEKO3


Powered by WEKO3