{"_buckets": {"deposit": "64893552-869f-4915-a769-37ec1240cc85"}, "_deposit": {"created_by": 1, "id": "71751", "owners": [1], "pid": {"revision_id": 0, "type": "depid", "value": "71751"}, "status": "published"}, "_oai": {"id": "oai:repo.qst.go.jp:00071751", "sets": ["28"]}, "author_link": ["706226", "706228", "706237", "706235", "706229", "706236", "706233", "706234", "706231", "706227", "706232", "706230"], "item_10005_date_7": {"attribute_name": "発表年月日", "attribute_value_mlt": [{"subitem_date_issued_datetime": "2014-09-17", "subitem_date_issued_type": "Issued"}]}, "item_10005_description_5": {"attribute_name": "抄録", "attribute_value_mlt": [{"subitem_description": "It is essential to consider large-angle scattered particles in dose calculation models for therapeutic ion beams. However, it is difficult to measure the small dose contribution from large-angle scattered particles. Therefore, we developed a parallel-plate ionization chamber consisting of concentric electrodes (ICCE) to efficiently and easily detect small contributions. The ICCE consists of two successive ICs with a common HV plate. The former is a large plane-parallel IC to measure dose distribution integrated over the whole plane, the latter is a 24-channel parallel-plate IC with concentric electrodes to derive the characteristic parameters describing the lateral beam spread. The aim of this study is to evaluate the performance of the ICCE. By taking advantage of the characteristic of ICCE, we studied the recombination associated with lateral beam profile. Also, we measured carbon pencil beam in several different media by using ICCE. As a result, we confirmed the ICCE could be used as a useful tool to determine the characterization of the therapeutic ion beams.", "subitem_description_type": "Abstract"}]}, "item_10005_description_6": {"attribute_name": "会議概要(会議名, 開催地, 会期, 主催者等)", "attribute_value_mlt": [{"subitem_description": "The 3rd International Beam Instrumentation Coference (IBIC2014)", "subitem_description_type": "Other"}]}, "item_access_right": {"attribute_name": "アクセス権", "attribute_value_mlt": [{"subitem_access_right": "metadata only access", "subitem_access_right_uri": "http://purl.org/coar/access_right/c_14cb"}]}, "item_creator": {"attribute_name": "著者", "attribute_type": "creator", "attribute_value_mlt": [{"creatorNames": [{"creatorName": "Hara, Yousuke"}], "nameIdentifiers": [{"nameIdentifier": "706226", "nameIdentifierScheme": "WEKO"}]}, {"creatorNames": [{"creatorName": "Furukawa, Takuji"}], "nameIdentifiers": [{"nameIdentifier": "706227", "nameIdentifierScheme": "WEKO"}]}, {"creatorNames": [{"creatorName": "Mizushima, Kota"}], "nameIdentifiers": [{"nameIdentifier": "706228", "nameIdentifierScheme": "WEKO"}]}, {"creatorNames": [{"creatorName": "Noda, Koji"}], "nameIdentifiers": [{"nameIdentifier": "706229", "nameIdentifierScheme": "WEKO"}]}, {"creatorNames": [{"creatorName": "Shirai, Toshiyuki"}], "nameIdentifiers": [{"nameIdentifier": "706230", "nameIdentifierScheme": "WEKO"}]}, {"creatorNames": [{"creatorName": "Tansho, Ryohei"}], "nameIdentifiers": [{"nameIdentifier": "706231", "nameIdentifierScheme": "WEKO"}]}, {"creatorNames": [{"creatorName": "原 洋介", "creatorNameLang": "en"}], "nameIdentifiers": [{"nameIdentifier": "706232", "nameIdentifierScheme": "WEKO"}]}, {"creatorNames": [{"creatorName": "古川 卓司", "creatorNameLang": "en"}], "nameIdentifiers": [{"nameIdentifier": "706233", "nameIdentifierScheme": "WEKO"}]}, {"creatorNames": [{"creatorName": "水島 康太", "creatorNameLang": "en"}], "nameIdentifiers": [{"nameIdentifier": "706234", "nameIdentifierScheme": "WEKO"}]}, {"creatorNames": [{"creatorName": "野田 耕司", "creatorNameLang": "en"}], "nameIdentifiers": [{"nameIdentifier": "706235", "nameIdentifierScheme": "WEKO"}]}, {"creatorNames": [{"creatorName": "白井 敏之", "creatorNameLang": "en"}], "nameIdentifiers": [{"nameIdentifier": "706236", "nameIdentifierScheme": "WEKO"}]}, {"creatorNames": [{"creatorName": "丹正 亮平", "creatorNameLang": "en"}], "nameIdentifiers": [{"nameIdentifier": "706237", "nameIdentifierScheme": "WEKO"}]}]}, "item_language": {"attribute_name": "言語", "attribute_value_mlt": [{"subitem_language": "eng"}]}, "item_resource_type": {"attribute_name": "資源タイプ", "attribute_value_mlt": [{"resourcetype": "conference object", "resourceuri": "http://purl.org/coar/resource_type/c_c94f"}]}, "item_title": "Dosimetric verification of lateral profile with a unique ionization chamber in therapeutic ion", "item_titles": {"attribute_name": "タイトル", "attribute_value_mlt": [{"subitem_title": "Dosimetric verification of lateral profile with a unique ionization chamber in therapeutic ion"}]}, "item_type_id": "10005", "owner": "1", "path": ["28"], "permalink_uri": "https://repo.qst.go.jp/records/71751", "pubdate": {"attribute_name": "公開日", "attribute_value": "2015-06-30"}, "publish_date": "2015-06-30", "publish_status": "0", "recid": "71751", "relation": {}, "relation_version_is_last": true, "title": ["Dosimetric verification of lateral profile with a unique ionization chamber in therapeutic ion"], "weko_shared_id": -1}
Dosimetric verification of lateral profile with a unique ionization chamber in therapeutic ion
https://repo.qst.go.jp/records/71751
https://repo.qst.go.jp/records/717518cd309f8-a04f-491c-8101-f72366074f8a
Item type | 会議発表用資料 / Presentation(1) | |||||
---|---|---|---|---|---|---|
公開日 | 2015-06-30 | |||||
タイトル | ||||||
タイトル | Dosimetric verification of lateral profile with a unique ionization chamber in therapeutic ion | |||||
言語 | ||||||
言語 | eng | |||||
資源タイプ | ||||||
資源タイプ識別子 | http://purl.org/coar/resource_type/c_c94f | |||||
資源タイプ | conference object | |||||
アクセス権 | ||||||
アクセス権 | metadata only access | |||||
アクセス権URI | http://purl.org/coar/access_right/c_14cb | |||||
著者 |
Hara, Yousuke
× Hara, Yousuke× Furukawa, Takuji× Mizushima, Kota× Noda, Koji× Shirai, Toshiyuki× Tansho, Ryohei× 原 洋介× 古川 卓司× 水島 康太× 野田 耕司× 白井 敏之× 丹正 亮平 |
|||||
抄録 | ||||||
内容記述タイプ | Abstract | |||||
内容記述 | It is essential to consider large-angle scattered particles in dose calculation models for therapeutic ion beams. However, it is difficult to measure the small dose contribution from large-angle scattered particles. Therefore, we developed a parallel-plate ionization chamber consisting of concentric electrodes (ICCE) to efficiently and easily detect small contributions. The ICCE consists of two successive ICs with a common HV plate. The former is a large plane-parallel IC to measure dose distribution integrated over the whole plane, the latter is a 24-channel parallel-plate IC with concentric electrodes to derive the characteristic parameters describing the lateral beam spread. The aim of this study is to evaluate the performance of the ICCE. By taking advantage of the characteristic of ICCE, we studied the recombination associated with lateral beam profile. Also, we measured carbon pencil beam in several different media by using ICCE. As a result, we confirmed the ICCE could be used as a useful tool to determine the characterization of the therapeutic ion beams. | |||||
会議概要(会議名, 開催地, 会期, 主催者等) | ||||||
内容記述タイプ | Other | |||||
内容記述 | The 3rd International Beam Instrumentation Coference (IBIC2014) | |||||
発表年月日 | ||||||
日付 | 2014-09-17 | |||||
日付タイプ | Issued |