Monitor Unit Calculation for Broad Beam Carbon-ion Therapy

https://repo.qst.go.jp/records/71134
https://repo.qst.go.jp/records/71134
a4706e48-2aed-402e-ad31-ab14968804d1
Item type 会議発表用資料 / Presentation(1)
公開日 2013-06-10
タイトル
タイトル Monitor Unit Calculation for Broad Beam Carbon-ion Therapy
言語
言語 eng
資源タイプ
資源タイプ識別子 http://purl.org/coar/resource_type/c_c94f
資源タイプ conference object
アクセス権
アクセス権 metadata only access
アクセス権URI http://purl.org/coar/access_right/c_14cb
著者 Mizota, Manabu

× Mizota, Manabu

WEKO 699262

Mizota, Manabu

Search repository
溝田 学

× 溝田 学

WEKO 699263

en 溝田 学

Search repository
抄録
内容記述タイプ Abstract
内容記述 Aim:
For long periods of time, dose calibration measurements have been done for each therapeutic beam in HIMAC heavy ion radiotherapy.
These days the number of treatment is increasing, so we have developed a calculation method to deliver monitor units of high precision.
The computational derivation helps to avoid the heavy burden on medical staffs and errors from wrong measurement.
\nMethod:
In carbon beam dosimetry, monitor unit depends primarily on the 3-dimensional field condition of broad beam method and secondary on the aperture made by collimator.
The aperture varies with patients infinitely whereas the number of field conditions is limited.
Precise aperture effect calculation is done using kernel beam model which is composed of two Gaussian functions.
For each field condition, parameters of the Gaussians are determined by optimization to fit measurement results of some stereotyped apertures. In addition, the kernel beam is weighted according to the flatness of the broad beam, which is effective especially in case of off-center positional calculation.
\nResults:
This method can calculate the fine differences among the value from the various apertures in shape with the same area.
The verification has been conducted by inspection, where calculated values of calibration factor using this method have been compared with the measured ones used for therapy. They are all in good agreement.
\nConclusions:
The monitor unit for broad beam carbon-ion therapy can be calculated in high precision, even in a small aperture. We have put it into practice since November, 2012.
会議概要(会議名, 開催地, 会期, 主催者等)
内容記述タイプ Other
内容記述 52nd Annual Conference of the Particle Therapy Co-Operative Group
発表年月日
日付 2013-06-08
日付タイプ Issued
戻る
0
views
See details
Views

Versions

Ver.1 2023-05-15 19:54:48.454830
Show All versions

Share

Mendeley Twitter Facebook Print Addthis

Cite as

エクスポート

OAI-PMH
  • OAI-PMH JPCOAR 2.0
  • OAI-PMH JPCOAR 1.0
  • OAI-PMH DublinCore
  • OAI-PMH DDI
Other Formats
  • JSON
  • BIBTEX

Confirm


Powered by WEKO3


Powered by WEKO3