Correction of head movement by frame-to-frame image realignment on human brain PET images with [11C]raclopride and [11C]FLB457

https://repo.qst.go.jp/records/70849
https://repo.qst.go.jp/records/70849
2760974b-95df-4f4d-bdf2-7fb1328c97aa
Item type 会議発表用資料 / Presentation(1)
公開日 2012-08-15
タイトル
タイトル Correction of head movement by frame-to-frame image realignment on human brain PET images with [11C]raclopride and [11C]FLB457
言語
言語 eng
資源タイプ
資源タイプ識別子 http://purl.org/coar/resource_type/c_c94f
資源タイプ conference object
アクセス権
アクセス権 metadata only access
アクセス権URI http://purl.org/coar/access_right/c_14cb
著者 Ikoma, Youko

× Ikoma, Youko

WEKO 696184

Ikoma, Youko

Search repository
Kimura, Yasuyuki

× Kimura, Yasuyuki

WEKO 696185

Kimura, Yasuyuki

Search repository
Shiraishi, Takahiro

× Shiraishi, Takahiro

WEKO 696186

Shiraishi, Takahiro

Search repository
Suhara, Tetsuya

× Suhara, Tetsuya

WEKO 696187

Suhara, Tetsuya

Search repository
Ito, Hiroshi

× Ito, Hiroshi

WEKO 696188

Ito, Hiroshi

Search repository
生駒 洋子

× 生駒 洋子

WEKO 696189

en 生駒 洋子

Search repository
木村 泰之

× 木村 泰之

WEKO 696190

en 木村 泰之

Search repository
白石 貴博

× 白石 貴博

WEKO 696191

en 白石 貴博

Search repository
須原 哲也

× 須原 哲也

WEKO 696192

en 須原 哲也

Search repository
伊藤 浩

× 伊藤 浩

WEKO 696193

en 伊藤 浩

Search repository
抄録
内容記述タイプ Abstract
内容記述 Objectives: Positron emission tomography with [11C]raclopride and [11C]FLB457 has been utilized for imaging dopamine D2 receptors in the striatum and extrastriatum, respectively. PET scans with these ligands require 60-90 min consecutive acquisition for a quantitative analysis. Therefore, head movement is often observed during scanning, and it hampers the reliability of quantitative outcomes, especially for small brain structures. Image-based motion correction with frame-to-frame realignment is a practical method, since it dose not require motion information obtained by an online tracking system. However, frame-to-frame realignment for emission PET images causes a mismatch between the emission scan and transmission scan used for the attenuation correction, and it may result in the error of quantitative analysis. In the present study, optimal image-based motion correction method including the transmission scan was evaluated for PET studies with [11C]raclopride and [11C]FLB457, and investigated the effect of this correction method on quantitative analysis outcomes.
Methods: In 60-min dynamic PET scan with [11C]raclopride and 90-min dynamic PET scan with [11C]FLB457, emission sinograms were reconstructed with or without attenuation correction. In the first method, 6 motion parameters (x-translation, y-translation, z-translation, roll, pitch, and yaw) were estimated frame-by-frame by realigning each frame of attenuation corrected image (AC) with a reference image using normalized mutual information [1]. Four kinds of images were used as the reference image for realignment, those are, summed image of all frames, summed image of the first 2 min, one frame image with high count, and MR image coregistered to PET summation image. Meanwhile, in the second method, motion parameters were estimated by realigning each frame of non-attenuation corrected images (NAC) by a similar way. Then µ-map obtained from a transmission scan was realigned using estimated motion parameters to correct a mismatch between the transmission and emission scans [2], and attenuation correction and reconstruction were performed sequentially. After motion correction, time-activity curves (TAC) for the striatum or frontal cortex were obtained. The binding potentials (BP) were estimated for these TACs by a simplified reference tissue model with the cerebellum as a reference region, and compared with BP estimates before the correction.
Results: The motion correction using NAC images could realign each frame image to every reference images for both [11C]raclopride and [11C]FLB457 studies, and the mismatch between the emission and transmission scan was also corrected using estimated motion parameters. However, the motion correction using AC images did not realign each frame image. After the correction with NAC, a discontinuity of TACs in the striatum and frontal cortex was improved. In the subject with the large movement during PET scanning, BP value of the striatum increased about 25% after the motion correction.
Conclusion: Head movement during the PET dynamic scan with [11C]raclopride and [11C]FLB457 could be accurately corrected by applying the image-based realignment to non-attenuation corrected images.
会議概要(会議名, 開催地, 会期, 主催者等)
内容記述タイプ Other
内容記述 The 9th International Symposium on Functional Neuroreceptor Mapping of the Living Brain (NRM2012)
発表年月日
日付 2012-08-11
日付タイプ Issued
戻る
0
views
See details
Views

Versions

Ver.1 2023-05-15 19:58:06.973063
Show All versions

Share

Mendeley Twitter Facebook Print Addthis

Cite as

エクスポート

OAI-PMH
  • OAI-PMH JPCOAR 2.0
  • OAI-PMH JPCOAR 1.0
  • OAI-PMH DublinCore
  • OAI-PMH DDI
Other Formats
  • JSON
  • BIBTEX

Confirm


Powered by WEKO3


Powered by WEKO3