Performance evaluation of four-layer DOI detectors

https://repo.qst.go.jp/records/70337
https://repo.qst.go.jp/records/70337
c8e077d8-6ee3-4a35-8b14-e9d5f5be133a
Item type 会議発表用資料 / Presentation(1)
公開日 2010-11-24
タイトル
タイトル Performance evaluation of four-layer DOI detectors
言語
言語 eng
資源タイプ
資源タイプ識別子 http://purl.org/coar/resource_type/c_c94f
資源タイプ conference object
アクセス権
アクセス権 metadata only access
アクセス権URI http://purl.org/coar/access_right/c_14cb
著者 Nishikido, Fumihiko

× Nishikido, Fumihiko

WEKO 690714

Nishikido, Fumihiko

Search repository
et.al

× et.al

WEKO 690715

et.al

Search repository
錦戸 文彦

× 錦戸 文彦

WEKO 690716

en 錦戸 文彦

Search repository
抄録
内容記述タイプ Abstract
内容記述 Multi pixel photon counters (MPPC) have many advantages to be used in PET detector, such as high internal gain, high photon detection efficiency, lower power supply and insensitivity to magnetic fields. However, the saturation effect is unavoidable phenomenon for PET detectors using MPPCs which need to collect large number of scintillation photons to achieve high resolutions. In a lot of cases, pulse height of PET detectors using MPPCs are saturated in high energy region. On the other hand, since our DOI encoding method previously presented is sharing scintillation photons to multiple pixels, the saturation effect can be suppressed. We constructed prototype four-layer detectors using two MPPC arrays and evaluated their detector performances. We constructed the prototype detector consisting of four layers of a 6 x 6 array with Lu2(1-x)Y2x SiO5 (LYSO) (Lu: 98 %, Y: 2 %) (Proteus Inc., U. S. A.) crystals, and a MPPC array with a 4 x 4 pixels array (S11064 series, Hamamatsu Photonics K.K. ). The size of each crystal element is 1.46 mm x 1.46 mm x 4.5 mm and all surfaces of crystal elements were chemically etched. For comparison, the two types of MPPC, S11064-050P (3600 cells) and S11064-025P (14400 cells), were used in the experiment. As a result, the all of the crystals of the detector using 050P were clearly identified while the crystal identification performance of the detector using 025P was degraded due to lower photon detection efficiency. The energy resolution of 11.0-13.6% (050P) and 12.7-15.2% (025P) were achieved for each detector. No saturations were observed for both detectors under the 511keV energy region.
会議概要(会議名, 開催地, 会期, 主催者等)
内容記述タイプ Other
内容記述 2010 Nuclear Science Symposium and Medical Imaging Conference
発表年月日
日付 2010-11-06
日付タイプ Issued
戻る
0
views
See details
Views

Versions

Ver.1 2023-05-15 20:04:06.343816
Show All versions

Share

Mendeley Twitter Facebook Print Addthis

Cite as

エクスポート

OAI-PMH
  • OAI-PMH JPCOAR 2.0
  • OAI-PMH JPCOAR 1.0
  • OAI-PMH DublinCore
  • OAI-PMH DDI
Other Formats
  • JSON
  • BIBTEX

Confirm


Powered by WEKO3


Powered by WEKO3