Calibration of Patient Positioning System using FPD at HIMAC

https://repo.qst.go.jp/records/70104
https://repo.qst.go.jp/records/70104
a8c04a6d-812d-437f-a67f-f580e0d4f0a1
Item type 会議発表用資料 / Presentation(1)
公開日 2010-05-25
タイトル
タイトル Calibration of Patient Positioning System using FPD at HIMAC
言語
言語 eng
資源タイプ
資源タイプ識別子 http://purl.org/coar/resource_type/c_c94f
資源タイプ conference object
アクセス権
アクセス権 metadata only access
アクセス権URI http://purl.org/coar/access_right/c_14cb
著者 Nitta, Kazunori

× Nitta, Kazunori

WEKO 688406

Nitta, Kazunori

Search repository
et.al

× et.al

WEKO 688407

et.al

Search repository
新田 和範

× 新田 和範

WEKO 688408

en 新田 和範

Search repository
抄録
内容記述タイプ Abstract
内容記述 Patient positioning is one of important procedures in radiation therapy because it is necessary to position patients properly according to their treatment planning. Flat Panel Detectors (FPDs) was installed for patient position detection system using X-ray in place of image instensifier tubes at National Institute of Radiological Sciences in August, 2009. FPD has many advantages compared with Image Intensifier tube; a) no image distortion, b) higher position resolution, c) wider dinamic range and d) decrease in halation effect due to density differnece. I report the installation of new positoning system including FPD, softwares and networking. I also show QC/QA procedures for FPDs.
After installing this system, I measured FPD performance and calibrated FPD position and pixel spacing which means the distance beteween pixels at iso-center. To measure the pixel spacing, I tried three kinds of methods (calculating geometrically, pointing manually two points on a grid board between which distance is known, and analyzing automatically image of grid by computer.)
Values of pixel spacing by three methods are all consistent. I summarized checking list for QA. Now new paitient positioning system is normally in operation. Whenever problem about FPD happens, I work to settle it immidiatelly.
Now new positioning system is working very well. FPD degradation due to the irradiation beam will be studied.
会議概要(会議名, 開催地, 会期, 主催者等)
内容記述タイプ Other
内容記述 PTCOG49
発表年月日
日付 2010-05-22
日付タイプ Issued
戻る
0
views
See details
Views

Versions

Ver.1 2023-05-15 20:06:47.330961
Show All versions

Share

Mendeley Twitter Facebook Print Addthis

Cite as

エクスポート

OAI-PMH
  • OAI-PMH JPCOAR 2.0
  • OAI-PMH JPCOAR 1.0
  • OAI-PMH DublinCore
  • OAI-PMH DDI
Other Formats
  • JSON
  • BIBTEX

Confirm


Powered by WEKO3


Powered by WEKO3