WEKO3
アイテム
Feasibility study of entire whole-body PET scanners based on the OpenPET geometry
https://repo.qst.go.jp/records/69929
https://repo.qst.go.jp/records/699295a2a8dcd-e93b-4642-87e3-13014e1ef279
Item type | 会議発表用資料 / Presentation(1) | |||||
---|---|---|---|---|---|---|
公開日 | 2009-11-09 | |||||
タイトル | ||||||
タイトル | Feasibility study of entire whole-body PET scanners based on the OpenPET geometry | |||||
言語 | ||||||
言語 | eng | |||||
資源タイプ | ||||||
資源タイプ識別子 | http://purl.org/coar/resource_type/c_c94f | |||||
資源タイプ | conference object | |||||
アクセス権 | ||||||
アクセス権 | metadata only access | |||||
アクセス権URI | http://purl.org/coar/access_right/c_14cb | |||||
著者 |
Yoshida, Eiji
× Yoshida, Eiji× et.al× 吉田 英治 |
|||||
抄録 | ||||||
内容記述タイプ | Abstract | |||||
内容記述 | A conventional PET scanner has a 15-25 cm axial field-of-view (FOV) and images a whole body using about six bed positions. An OpenPET geometry can extend the axial FOV with a limited number of detectors. The entire whole-body PET scanner must be able to process a huge amount of data, effectively. In this work, we study feasibility of the fully 3D entire whole-body PET scanner using the GATE simulation. The OpenPET has 12 block detector rings with the ring diameter of 840 mm and each block detector ring consists of 48 DOI (depth-of-interaction) detectors. The OpenPET has the axial length of 895.95 mm with five parts of 58.95 mm open gaps. The OpenPET has higher single data loss than a conventional PET scanner at grouping circuits. NECR of the OpenPET decreases by single data loss. But, single data loss is improved by separating the axially arranged detector into two parts. Also, multiple coincidences are found to be important for the entire whole-body PET scanner. The entire whole-body PET scanner with the OpenPET geometry promises to provide a large axial FOV with the open space and to have sufficient performance values. But, single data loss at the grouping circuit and multiple coincidences are limited to the peak NECR for the entire whole-body PET canner. | |||||
会議概要(会議名, 開催地, 会期, 主催者等) | ||||||
内容記述タイプ | Other | |||||
内容記述 | Nuclear Science Symposium and Medical Imaging Conference | |||||
発表年月日 | ||||||
日付 | 2009-10-31 | |||||
日付タイプ | Issued |