A MATHEMATICAL MORPHOLOGY BASED METAPHASE FINDER FOR THE BIOLOGICAL DOSIMETRY BY CHROMOSOME ABERRATIONS

https://repo.qst.go.jp/records/69461
https://repo.qst.go.jp/records/69461
2fa1f548-81a7-4af1-b6c4-d008d6060635
Item type 会議発表用資料 / Presentation(1)
公開日 2008-09-19
タイトル
タイトル A MATHEMATICAL MORPHOLOGY BASED METAPHASE FINDER FOR THE BIOLOGICAL DOSIMETRY BY CHROMOSOME ABERRATIONS
言語
言語 eng
資源タイプ
資源タイプ識別子 http://purl.org/coar/resource_type/c_c94f
資源タイプ conference object
アクセス権
アクセス権 metadata only access
アクセス権URI http://purl.org/coar/access_right/c_14cb
著者 Furukawa, Akira

× Furukawa, Akira

WEKO 681698

Furukawa, Akira

Search repository
古川 章

× 古川 章

WEKO 681699

en 古川 章

Search repository
抄録
内容記述タイプ Abstract
内容記述 For biological dosimetry by counting chromosome aberrations, automation technique has been required to process large number of sample preparations at low dose radiation exposure. The metaphase finder is an automated optical microscope system, which automatically scans and finds metaphase cells on the slide glass, and relocates metaphase cells to the center of the field of view of the microscope to observe chromosomes in high magnification.
We have constructed a cost-effective metaphase finder system by assembling commercially available components, such as microscope, motorized sample stage, personal computer and general purpose image analysis software package, instead of special dedicated system which we have made previously (1). The new system has high flexibility in adapting to the new staining methods.
We have utilized the mathematical morphology based image processing to the algorithm to find metaphase cells. Especially, the morphology for the gray-image was used for the newest version.
We tested the performance of this metaphase finder system. The scanning speed was 18 min/cm2. The recent improvement enabled metaphase finding in the center and peripherals of the images which were differently illuminated.
We have distributed this metaphase finder systems to 6 laboratories nationwide in Japan, which are known as The Chromosome Network for Biodosimetry in Japan (2), for the testing the new algorithm for the practical use.
会議概要(会議名, 開催地, 会期, 主催者等)
内容記述タイプ Other
内容記述 BioDose2008
発表年月日
日付 2008-09-10
日付タイプ Issued
戻る
0
views
See details
Views

Versions

Ver.1 2023-05-15 20:14:12.965570
Show All versions

Share

Mendeley Twitter Facebook Print Addthis

Cite as

エクスポート

OAI-PMH
  • OAI-PMH JPCOAR 2.0
  • OAI-PMH JPCOAR 1.0
  • OAI-PMH DublinCore
  • OAI-PMH DDI
Other Formats
  • JSON
  • BIBTEX

Confirm


Powered by WEKO3


Powered by WEKO3