WEKO3
アイテム
DEVELOPMNT OF 3D Does Verification System for Scanned Ion Beam AT HIMAC
https://repo.qst.go.jp/records/69400
https://repo.qst.go.jp/records/6940058149158-b053-440d-99f0-b1977adfef74
Item type | 会議発表用資料 / Presentation(1) | |||||
---|---|---|---|---|---|---|
公開日 | 2008-07-07 | |||||
タイトル | ||||||
タイトル | DEVELOPMNT OF 3D Does Verification System for Scanned Ion Beam AT HIMAC | |||||
言語 | ||||||
言語 | eng | |||||
資源タイプ | ||||||
資源タイプ識別子 | http://purl.org/coar/resource_type/c_c94f | |||||
資源タイプ | conference object | |||||
アクセス権 | ||||||
アクセス権 | metadata only access | |||||
アクセス権URI | http://purl.org/coar/access_right/c_14cb | |||||
著者 |
Saotome, Naoya
× Saotome, Naoya× 早乙女 直也 |
|||||
抄録 | ||||||
内容記述タイプ | Abstract | |||||
内容記述 | A 3D dose imaging system has been developedfor a project of a new cancer treatement with 3D pencil beam scanning at HIMAC. This system provides the dose measurements easily and rapidly. This system consists of a water tank, fluoresecent screen and charge-coupled device, set at isocentor. The fluorescent screen is directlly attached to the downstream side of water tank. One of great advantages of this system is to obtain 2D dose map at once, by correcting LET-dependent quenching. The proced to verify 3D dose distribution is based on the 2D dose measurement of slice-by-slice under a water depth. We will present the measurement result of 3D dose distribution by the prposed method ,and its comparison withthat by theionization chamber. | |||||
会議概要(会議名, 開催地, 会期, 主催者等) | ||||||
内容記述タイプ | Other | |||||
内容記述 | EPAC 2008 (11th European Particle Accelerator Conference) | |||||
発表年月日 | ||||||
日付 | 2008-06-27 | |||||
日付タイプ | Issued |