WEKO3
アイテム
Development of a Portable in vivo Monitor for Radiation Emergency Accidents
https://repo.qst.go.jp/records/68805
https://repo.qst.go.jp/records/6880574283d45-dd3e-43af-9e06-56dd2b0e8dda
| Item type | 会議発表用資料 / Presentation(1) | |||||
|---|---|---|---|---|---|---|
| 公開日 | 2006-11-30 | |||||
| タイトル | ||||||
| タイトル | Development of a Portable in vivo Monitor for Radiation Emergency Accidents | |||||
| 言語 | ||||||
| 言語 | eng | |||||
| 資源タイプ | ||||||
| 資源タイプ識別子 | http://purl.org/coar/resource_type/c_c94f | |||||
| 資源タイプ | conference object | |||||
| アクセス権 | ||||||
| アクセス権 | metadata only access | |||||
| アクセス権URI | http://purl.org/coar/access_right/c_14cb | |||||
| 著者 |
Eunjoo, Kim
× Eunjoo, Kim× Suzuki, Tosikazu× 金 ウンジュ× 鈴木 敏和 |
|||||
| 抄録 | ||||||
| 内容記述タイプ | Abstract | |||||
| 内容記述 | Our purpose in this study is to develop an in-vivo monitor with compact and portable which can be used on the scene. This monitor, which consists of a Ge detector and a guard detector, can identify and quantify the nuclei by applying the anti-coincidence method in spite of considerable background. In order to determine the Germanium crystal size and optimize a dewar size of a Ge detector, the fundamental property data of cooling system is needed in the design. In the present study, a characteristic test for Ge detectors has been carried out to examine the property of cooling system, such as the duration for which the temperature of Germanium crystal reached below -150 degrees and LN2 consumption. From the obtained results, it was found that the Ge detector will be used for responding to radiation emergency accidents. | |||||
| 発表年月日 | ||||||
| 日付 | 2005-07-28 | |||||
| 日付タイプ | Issued | |||||