Preliminary Performance Evaluation of the Prototype System for a Brain DOI-PET Scanner: jPET-D4
https://repo.qst.go.jp/records/67915
https://repo.qst.go.jp/records/67915b87d7af2-aae3-4ea7-bf0b-f95b3b942f05
Item type | 会議発表用資料 / Presentation(1) | |||||
---|---|---|---|---|---|---|
公開日 | 2004-10-27 | |||||
タイトル | ||||||
タイトル | Preliminary Performance Evaluation of the Prototype System for a Brain DOI-PET Scanner: jPET-D4 | |||||
言語 | ||||||
言語 | eng | |||||
資源タイプ | ||||||
資源タイプ識別子 | http://purl.org/coar/resource_type/c_c94f | |||||
資源タイプ | conference object | |||||
アクセス権 | ||||||
アクセス権 | metadata only access | |||||
アクセス権URI | http://purl.org/coar/access_right/c_14cb | |||||
著者 |
Yamaya, Taiga
× Yamaya, Taiga× et.al× 山谷 泰賀 |
|||||
抄録 | ||||||
内容記述タイプ | Abstract | |||||
内容記述 | A high-performance brain PET scanner, jPET-D4, which provides depth-of-interaction (DOI) information, is under development. The scanner is designed to achieve not only high spatial resolution but also high scanner sensitivity with the DOI information obtained from multi-layered thin crystals. The scanner has 5 rings of 24 detector blocks each, and the detector block consists of 1,024 GSO crystals of 2.9mm x 2.9mm x 7.5mm, which are arranged in 4 layers of 16 x 16 arrays. At this stage, a pair of detector blocks and a coincidence circuit has been assembled into an experimental prototype gantry. In this paper, as a preliminary experiment, we have investigated the performance of the jPET-D4's inherent spatial resolution and noise characteristics using the prototype system. Spatial resolution and background noise were measured by using a line source and a uniform cylinder phantom respectively. To avoid systematic error and reduce computational cost in image reconstruction, the DOI compression (DOIC) method followed by the ML-EM with accurate system modeling, that we have previously proposed, was applied. Experimental results showed that the jPET-D4 achieves less than 3mm spatial resolution over the field-of-view. Trade-off results between spatial resolution and background noise also promise an excellent performance of the scanner. | |||||
会議概要(会議名, 開催地, 会期, 主催者等) | ||||||
内容記述タイプ | Other | |||||
内容記述 | IEEE 2004 Nuclear Science Symposium and Medical Imaging Conference | |||||
発表年月日 | ||||||
日付 | 2004-10-22 | |||||
日付タイプ | Issued |