RF SYSTEM FOR COMPACT MEDICAL PROTON SYNCHROTRON
https://repo.qst.go.jp/records/67843
https://repo.qst.go.jp/records/678431f8cc5ba-447f-487c-99fd-ae3976ed2fa7
Item type | 会議発表用資料 / Presentation(1) | |||||
---|---|---|---|---|---|---|
公開日 | 2004-07-27 | |||||
タイトル | ||||||
タイトル | RF SYSTEM FOR COMPACT MEDICAL PROTON SYNCHROTRON | |||||
言語 | ||||||
言語 | eng | |||||
資源タイプ | ||||||
資源タイプ識別子 | http://purl.org/coar/resource_type/c_c94f | |||||
資源タイプ | conference object | |||||
アクセス権 | ||||||
アクセス権 | metadata only access | |||||
アクセス権URI | http://purl.org/coar/access_right/c_14cb | |||||
著者 |
Fan
× Fan× 方 志高 |
|||||
抄録 | ||||||
内容記述タイプ | Abstract | |||||
内容記述 | The rf system has been developed for the compact medical proton synchrotron. The rf cavity consists of 2 cells loaded with 4 high-permeability magnetic alloy cores in each cell, and the 2 cells of rf cavity are driven by a push-pull power amplifier in parallel which is formed with two tetrode tubes 4CX35,000C. The rf system is operated in pulse mode. During the beam acceleration time of 5 ms, the rf frequency sweeps from 2.0MHz to 17.8MHz and the cavity voltage varies from 10.0kV to 1.6kV. The high power test of the rf system is being performed, and the required gap voltage has been successfully achieved. The performance of the rf system will be presented. | |||||