WEKO3
アイテム
Estimation of a Clinical Dose of Heavy-Ion Radiotherapy Using a Tissue Equivalent Proportional Counter
https://repo.qst.go.jp/records/67458
https://repo.qst.go.jp/records/67458fc4501db-9032-49c9-92ac-9c5f305a77aa
Item type | 会議発表用資料 / Presentation(1) | |||||
---|---|---|---|---|---|---|
公開日 | 2003-09-03 | |||||
タイトル | ||||||
タイトル | Estimation of a Clinical Dose of Heavy-Ion Radiotherapy Using a Tissue Equivalent Proportional Counter | |||||
言語 | ||||||
言語 | eng | |||||
資源タイプ | ||||||
資源タイプ識別子 | http://purl.org/coar/resource_type/c_c94f | |||||
資源タイプ | conference object | |||||
アクセス権 | ||||||
アクセス権 | metadata only access | |||||
アクセス権URI | http://purl.org/coar/access_right/c_14cb | |||||
著者 |
Kase, Yuuki
× Kase, Yuuki× 加瀬 優紀 |
|||||
抄録 | ||||||
内容記述タイプ | Abstract | |||||
内容記述 | Clinical dose of heavy-ion therapy at HIMAC is defined by physical dose multiplied by RBE. The RBE of heavy ion beam is calculated using LQ model based on experimental survival curve for mono-energetic beams. From this definition, it is very hard to obtain RBE value for heavy ion fields which is not standard mixture of the mono-energetic beams. The aim of this work is to establish the method to estimate the clinical dose easily with a tissue equivalent proportional counter. The results for carbon beam of 290MeV/n and 60mm SOBP are reported here. |
|||||
会議概要(会議名, 開催地, 会期, 主催者等) | ||||||
内容記述タイプ | Other | |||||
内容記述 | World Congress on Medical Physics and Biomedical Engineering 2003 | |||||
発表年月日 | ||||||
日付 | 2003-08-29 | |||||
日付タイプ | Issued |