Validation of radon-film-badge for measurement of radon in water

https://repo.qst.go.jp/records/66932
https://repo.qst.go.jp/records/66932
e9272e75-b67a-4244-8ba0-cf85b33dad0e
Item type 会議発表用資料 / Presentation(1)
公開日 2018-09-28
タイトル
タイトル Validation of radon-film-badge for measurement of radon in water
言語
言語 eng
資源タイプ
資源タイプ識別子 http://purl.org/coar/resource_type/c_c94f
資源タイプ conference object
アクセス権
アクセス権 metadata only access
アクセス権URI http://purl.org/coar/access_right/c_14cb
著者 ミロソラフ, ヤニック

× ミロソラフ, ヤニック

WEKO 657954

ミロソラフ, ヤニック

Search repository
Yasuoka, Y.

× Yasuoka, Y.

WEKO 657955

Yasuoka, Y.

Search repository
Higuchi, S.

× Higuchi, S.

WEKO 657956

Higuchi, S.

Search repository
M., Hasan Md.

× M., Hasan Md.

WEKO 657957

M., Hasan Md.

Search repository
Tommasino, L.

× Tommasino, L.

WEKO 657958

Tommasino, L.

Search repository
ミロソラフ ヤニック

× ミロソラフ ヤニック

WEKO 657959

en ミロソラフ ヤニック

Search repository
抄録
内容記述タイプ Abstract
内容記述 Ground and surface water usually contains a certain amount of radionuclides, including radon, which dissolves from rocks and soil into water.
The current methods for radon measurement in water are mainly based on active measurement techniques, e.g. the degassing of radon using bubbling in combination with alpha or gamma spectrometry (bubbling method), gamma-ray spectrometry or liquid scintillation counting (LSC). These methods have some disadvantages that include poor resolution (gamma detectors), the influence of water temperature and the level of degassing during transport (bubbling method).
Therefore, a passive, cheap and effective method based on CR-39 chips may be a good candidate for wide-ranging of surveys where the transportation of water samples would be troublesome. This method could be used not only for radon protection surveys, but also for geo-hazard research.
The present paper describes the validation of the recently developed radon-film-badge for radon measurements in water using RAD7 and LSC methods.
A radon-film-badge was constructed incorporating a thin-film radiator (Makrofol-polycarbonate) with suitable radon-sorption characteristics, placed next to a track detector (CR-39), then sealed a polyethylene bag.
Different radon-badges were exposed to highly radon contaminated water from a 70 m deep well located in Kobe. The same water was measured using an electrostatic type radon monitor (RAD7) in 250 ml vials, and with the liquid scintillation method (LSC).
In addition, the calibration factor for the radon-film-badge in air was obtained at the QST/NIRS radon facility.
会議概要(会議名, 開催地, 会期, 主催者等)
内容記述タイプ Other
内容記述 9th International Conference on High Level Environmental Radiation Areas - For Understanding Chronic Low-Dose-Rate Radiation Exposure Health Effects and Social Impacts (ICHLERA 2018)における発表
発表年月日
日付 2018-09-26
日付タイプ Issued
戻る
0
views
See details
Views