Basic characteristics of alumina-based ceramics TLD to charged particle beams

https://repo.qst.go.jp/records/65921
https://repo.qst.go.jp/records/65921
b5975ab3-b8b4-4634-bc60-6037a4167bd7
Item type 会議発表用資料 / Presentation(1)
公開日 2016-07-13
タイトル
タイトル Basic characteristics of alumina-based ceramics TLD to charged particle beams
言語
言語 jpn
資源タイプ
資源タイプ識別子 http://purl.org/coar/resource_type/c_c94f
資源タイプ conference object
アクセス権
アクセス権 metadata only access
アクセス権URI http://purl.org/coar/access_right/c_14cb
著者 古場, 裕介

× 古場, 裕介

WEKO 649360

古場, 裕介

Search repository
Chang, Weishan

× Chang, Weishan

WEKO 649361

Chang, Weishan

Search repository
Shinsho, Kiyomitsu

× Shinsho, Kiyomitsu

WEKO 649362

Shinsho, Kiyomitsu

Search repository
Yanagisawa, Shin

× Yanagisawa, Shin

WEKO 649363

Yanagisawa, Shin

Search repository
Sasaki, Daich

× Sasaki, Daich

WEKO 649364

Sasaki, Daich

Search repository
Wakabayashi, Genichiro

× Wakabayashi, Genichiro

WEKO 649365

Wakabayashi, Genichiro

Search repository
古場 裕介

× 古場 裕介

WEKO 649366

en 古場 裕介

Search repository
抄録
内容記述タイプ Abstract
内容記述 1. Introduction
Charged particle therapy, such as proton and carbon therapy, became popular internationally as a radiation modality for cancer treatment. The verifications of the irradiation field or dose distribution in the charged particle therapy require the measurements with high spatial resolution, because these distributions are precipitous and complex. Recently, it was reported that the commercial Al2O3 ceramics plate showed good TL characteristics and available as 2D dosimeter [1]. This ceramic TLD was very easy to deal and low-cost device, and with high spatial resolution. In order to use this ceramic TLD in the charged particle therapy as the verification tools, we investigated the thermoluminescence characteristics of the ceramic TLD to charged particle beams.
2. Materials and Methods
In this study, the ceramics TLDs (Chibacera cop. A8; Al2O3 : >99.5 wt%, SiO2 : <0.10 wt%, Fe2O3: <0.05 wt%, Na2O: <0.10 wt%) were used. Experiments of several charged-particle-beam irradiations were performed at the Heavy-Ion Medical Accelerator in Chiba (HIMAC) in the National Institute of Radiological Sciences. The beams of proton 160 MeV, He 150 MeV/u, C 290 MeV/u, Ne 400 MeV/u and Ar 500 MeV/u were used. And the X-ray irradiation was carried out using a Varian CLINAC-21EX Linear accelerator at 6 MV. In irradiations, four dose levels were given as 0.5, 1.0, 2.0 and 5.0 Gy in water concerning the predicted dose to the charged particle therapy. The glow curve of the ceramic TLD was measured using a photon counting unit (HAMAMATSU H11890-110) and a brass plate heater connected programmable heat controller (Sakaguchi E.H VOC CORP. SCR-SHQ-A). The glow curves were recorded from room temperature up to 400°C at a heating rate of 0.13°C/s in air. Thermoluminescence intensity was calibrated using 60Co gamma rays.
3. Results & Discussion
Figure 1 shows the glow curve of this alumina based ceramics TLD. The glow peaks were located at 148 °C (main peak) and 285 °C. Thermoluminescence intensity was evaluated using the range of 10 °C centered on the main peak. In any particle irradiation, dose linearity and repeatability was good comparatively. The standard deviation of variation coefficient of dose responses was less than 4% except for Ar irradiation. The TL efficiencies of each peak showed different LET dependence effect. Figure 2 shows the LET dependency of TL efficiency of main peak. The TL efficiency of the peak located at 148 °C showed small LET dependency as compared to common TLDs [2]. In the range from 0.1 to 20 keV/m, LET dependency was less than 10%. In Ar irradiation, TL efficiency was about 80% at 86 keV/m. In dosimetry of charged particle therapy, solid state detectors such as TLD were avoided to use generally, due to such large LET dependency. This ceramics TLD might be available as a verification tool in limited LET range like as proton therapy.
4. Conclusions
In this study, basic characteristics of alumina-based ceramics TLD to charged particle beams were reported. The glow curve had two peaks, and main peak located at 148 °C showed good linearity and repeatability. The TL efficiency of the main peak showed small LET dependency as compared to common TLDs. This ceramics TLD might be available as a verification tool in limited LET range like as proton therapy.
5. References
[1] K. Shinsho et al. X-ray imaging using thermoluminescence properties of commercial
Al2O3 ceramics plates, Applied Radiation and Isotopes, in press.
[2] T. Bergera, M. Hajek, TL-efficiency—Overview and experimental results over the years, Radiation Measurements 43 (2008) 146-156.
会議概要(会議名, 開催地, 会期, 主催者等)
内容記述タイプ Other
内容記述 The 18th International Conference on Solid State Dosimetry
発表年月日
日付 2016-07-05
日付タイプ Issued
戻る
0
views
See details
Views

Versions

Ver.1 2023-05-15 20:54:48.699252
Show All versions

Share

Mendeley Twitter Facebook Print Addthis

Cite as

エクスポート

OAI-PMH
  • OAI-PMH JPCOAR 2.0
  • OAI-PMH JPCOAR 1.0
  • OAI-PMH DublinCore
  • OAI-PMH DDI
Other Formats
  • JSON
  • BIBTEX

Confirm


Powered by WEKO3


Powered by WEKO3