| タイトル | ||||||
| タイトル | Feasibility study of novel geometry of an axially extendable multiplex cylinder PET | |||||
| 言語 | ||||||
| 言語 | eng | |||||
| 資源タイプ | ||||||
| 資源タイプ識別子 | http://purl.org/coar/resource_type/c_c94f | |||||
| 資源タイプ | conference output | |||||
| アクセス権 | ||||||
| アクセス権 | metadata only access | |||||
| アクセス権URI | http://purl.org/coar/access_right/c_14cb | |||||
| 著者 |
Yoshida, Eiji
× Yoshida, Eiji× Yamaya, Taiga× 吉田 英治× 山谷 泰賀 |
|||||
| 抄録 | ||||||
| 内容記述タイプ | Abstract | |||||
| 内容記述 | A current clinical PET scanner has a 15–22 cm axial field-of-view (FOV). These PET scanners image the whole body using about six bed positions. We proposed our original OpenPET geometry which can extend the axial FOV with a limited number of detectors. For an alternative approach, we then designed an axially extendable multiplex cylinder (AEMC) PET scanner to provide high versatility for clinical and research studies using a silicon-photomultiplier (Si-PM)-based depth-of-interaction (DOI) detector. Since Si-PMs have high gain like PMTs and a compact design, the Si-PM-based detector is expected to enable various new detector arrangements. The AEMC-PET scanner consists of four independent and laminated detector rings using a four-layer DOI detector. This PET scanner can extend the axial FOV as each stacked detector ring can be slid aside. When this PET scanner is used for the four-layer DOI detector, its minimum axial FOV is 24 cm and its maximum crystal thickness is 3 cm. On the other hand, the axial FOV can be extended to 96 cm while keeping a continuous detector ring, but the crystal thickness must be 1/4 of 3 cm. In addition, using the OpenPET geometry, the axial FOV can be extended to 168 cm. In this work, we studied the feasibility of the PET scanner with a variable axial FOV using Monte Carlo simulation. For the 180-cm line source simulation, the 96-cm axial FOV was obtained with twofold better sensitivity compared to the 24-cm axial FOV. Also, we showed that the AEMC-PET scanner had a continuous axial FOV to 168 cm. We expect the AEMC-PET scanner will provide high versatility such as for measuring trace whole-body uptakes while keeping the continuous axial FOV and the scan time for static images will be reduced with a limited number of detectors. | |||||
| 会議概要(会議名, 開催地, 会期, 主催者等) | ||||||
| 内容記述タイプ | Other | |||||
| 内容記述 | 2nd SNU-NIRS Workshop on Nuclear Medicine Imaging Science and Technology | |||||