Feasibility study of novel geometry of an axially extendable multiplex cylinder PET

https://repo.qst.go.jp/records/64990
https://repo.qst.go.jp/records/64990
0bcc5fb3-be0f-4be6-9e62-c1d65fe7ef55
アイテムタイプ 会議発表用資料 / Presentation(1)
公開日 2013-05-07
タイトル
タイトル Feasibility study of novel geometry of an axially extendable multiplex cylinder PET
言語
言語 eng
資源タイプ
資源タイプ識別子 http://purl.org/coar/resource_type/c_c94f
資源タイプ conference output
アクセス権
アクセス権 metadata only access
アクセス権URI http://purl.org/coar/access_right/c_14cb
著者 Yoshida, Eiji

× Yoshida, Eiji

WEKO 640378

Yoshida, Eiji

Search repository
Yamaya, Taiga

× Yamaya, Taiga

WEKO 640379

Yamaya, Taiga

Search repository
吉田 英治

× 吉田 英治

WEKO 640380

en 吉田 英治

Search repository
山谷 泰賀

× 山谷 泰賀

WEKO 640381

en 山谷 泰賀

Search repository
抄録
内容記述タイプ Abstract
内容記述 A current clinical PET scanner has a 15–22 cm axial field-of-view (FOV). These PET scanners image the whole body using about six bed positions. We proposed our original OpenPET geometry which can extend the axial FOV with a limited number of detectors. For an alternative approach, we then designed an axially extendable multiplex cylinder (AEMC) PET scanner to provide high versatility for clinical and research studies using a silicon-photomultiplier (Si-PM)-based depth-of-interaction (DOI) detector. Since Si-PMs have high gain like PMTs and a compact design, the Si-PM-based detector is expected to enable various new detector arrangements. The AEMC-PET scanner consists of four independent and laminated detector rings using a four-layer DOI detector. This PET scanner can extend the axial FOV as each stacked detector ring can be slid aside. When this PET scanner is used for the four-layer DOI detector, its minimum axial FOV is 24 cm and its maximum crystal thickness is 3 cm. On the other hand, the axial FOV can be extended to 96 cm while keeping a continuous detector ring, but the crystal thickness must be 1/4 of 3 cm. In addition, using the OpenPET geometry, the axial FOV can be extended to 168 cm. In this work, we studied the feasibility of the PET scanner with a variable axial FOV using Monte Carlo simulation. For the 180-cm line source simulation, the 96-cm axial FOV was obtained with twofold better sensitivity compared to the 24-cm axial FOV. Also, we showed that the AEMC-PET scanner had a continuous axial FOV to 168 cm. We expect the AEMC-PET scanner will provide high versatility such as for measuring trace whole-body uptakes while keeping the continuous axial FOV and the scan time for static images will be reduced with a limited number of detectors.
会議概要(会議名, 開催地, 会期, 主催者等)
内容記述タイプ Other
内容記述 2nd SNU-NIRS Workshop on Nuclear Medicine Imaging Science and Technology
発表年月日
日付 2013-04-25
日付タイプ Issued
戻る
0
views
See details
Views

Versions

Ver.1 2023-05-15 21:05:18.212428
Show All versions

Share

Share
tweet

Cite as

Other

print

エクスポート

OAI-PMH
  • OAI-PMH JPCOAR 2.0
  • OAI-PMH JPCOAR 1.0
  • OAI-PMH DublinCore
  • OAI-PMH DDI
Other Formats
  • JSON
  • BIBTEX
  • ZIP

コミュニティ

確認

確認

確認


Powered by WEKO3


Powered by WEKO3