Dose measurement for medical staff with glass dosimeters and thermoluminescence dosimeters during 125I brachytherapy

https://repo.qst.go.jp/records/63842
https://repo.qst.go.jp/records/63842
fb854291-2b11-4bfa-ae0f-7bbc99fd4c9c
Item type 会議発表用資料 / Presentation(1)
公開日 2010-03-15
タイトル
タイトル Dose measurement for medical staff with glass dosimeters and thermoluminescence dosimeters during 125I brachytherapy
言語
言語 eng
資源タイプ
資源タイプ識別子 http://purl.org/coar/resource_type/c_c94f
資源タイプ conference object
アクセス権
アクセス権 metadata only access
アクセス権URI http://purl.org/coar/access_right/c_14cb
著者 Fujii, Keisuke

× Fujii, Keisuke

WEKO 629831

Fujii, Keisuke

Search repository
Ko, Susumu

× Ko, Susumu

WEKO 629832

Ko, Susumu

Search repository
Nako, Yasunobu

× Nako, Yasunobu

WEKO 629833

Nako, Yasunobu

Search repository
Tonari, Ayako

× Tonari, Ayako

WEKO 629834

Tonari, Ayako

Search repository
Nishizawa, Kanae

× Nishizawa, Kanae

WEKO 629835

Nishizawa, Kanae

Search repository
Akahane, Keiichi

× Akahane, Keiichi

WEKO 629836

Akahane, Keiichi

Search repository
Takayama, Makoto

× Takayama, Makoto

WEKO 629837

Takayama, Makoto

Search repository
藤井 啓輔

× 藤井 啓輔

WEKO 629838

en 藤井 啓輔

Search repository
西澤 かな枝

× 西澤 かな枝

WEKO 629839

en 西澤 かな枝

Search repository
赤羽 恵一

× 赤羽 恵一

WEKO 629840

en 赤羽 恵一

Search repository
抄録
内容記述タイプ Abstract
内容記述 Brachytherapy for prostate cancer with 125I seeds is a well established procedure, and the number of this procedure has been increasing in Japan. While it was reported that this procedure showed high effectiveness, concerns arose over the radiation doses and safety for medical staff because medical staff, in particular, a surgeon would stand close to the source during this procedure. To ensure safety for medical staff, we measured surface doses to medical personnel during the 125I brachytherapy.
Dose measurements were performed with photoluminescent glass dosimeters (PLDs) and thermoluminescence dosimeters (TLDs). PLDs have the advantage of the large measurement dose range. PLDs are easy to handle in clinical environment since they can be read repeatedly. The PLD system consisted of silver-activated PLDs and a PL reader, and used pulsed UV laser stimulation. PLDs were used to measure surface doses to a surgeon and an assistant for 20 and 17 times, respectively. TLDs were used to check the accuracy of the measurement and to measure surface doses to a nurse, a radiologist, an anesthesiologist and a radiological technologist. These dosimeters were calibrated with a traceable ionization chamber for photon beam at the corresponding energy. The variation of sensitivity for PLDs and TLDs for the effective energy of 28-37 keV was estimated to be about 9% and 11%, respectively.
PLD readings agreed with those of TLDs within about 10% for doses above 100micro-Gy outside of a few exceptions. Surface doses at the surgeon's left arms and hand reached up to 509micro-Gy and 650micro-Gy, respectively. Doses at other monitoring points of a surgeon and at all points of an assistant were below 100micro-Gy. The doses at surgeon's hands in the lead gloves of 0.03mm Pb nominal thickness were 50-70% lower than those at arms out of gloves. The tube voltage, tube current and irradiation time in fluoroscopy decreased gradually due to the increase in the surgeon's experience. Doses at all monitoring points of a nurse, a radiologist, an anesthesiologist and a radiological technologist were below 10micro-Gy since they positioned away from the source, and they were shielded by a protective lead glass.
会議概要(会議名, 開催地, 会期, 主催者等)
内容記述タイプ Other
内容記述 European Conference on Individual Monitoring of Ionizing Radiation
発表年月日
日付 2010-03-12
日付タイプ Issued
戻る
0
views
See details
Views

Versions

Ver.1 2023-05-15 21:18:23.339905
Show All versions

Share

Mendeley Twitter Facebook Print Addthis

Cite as

エクスポート

OAI-PMH
  • OAI-PMH JPCOAR 2.0
  • OAI-PMH JPCOAR 1.0
  • OAI-PMH DublinCore
  • OAI-PMH DDI
Other Formats
  • JSON
  • BIBTEX

Confirm


Powered by WEKO3


Powered by WEKO3